Return to study ST001423 main page
MB Sample ID: SA119338
Local Sample ID: | 40002164_1 |
Subject ID: | SU001497 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Age Or Age Range: | 29-82 |
Gender: | Male and female |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002380 | AN002381 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | Reversed phase |
Chromatography system | Dionex UltiMate 3000 | Dionex UltiMate 3000 |
Column | Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge | Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE | POSITIVE |
Units | peak area | peak area |
Chromatography:
Chromatography ID: | CH001747 |
Chromatography Summary: | The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 microliters of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 2 min, increased to 0.4 mL/min at 6 min and held for 4 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 25% A, 70% B, 5% C hold for 2 min, with linear gradient to 75% A, 20% B, 5% C at 6 min, hold for 4 min, resulting in a total analytical run time of 10 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 25% A, 70% B, 5% C. |
Methods ID: | 2% formic acid in LC-MS grade water |
Methods Filename: | 20160120_posHILIC120kres10min_ESI_c18poswash_acrore.meth |
Chromatography Comments: | Triplicate injections for each chromatography mode |
Instrument Name: | Dionex UltiMate 3000 |
Column Name: | Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge |
Column Temperature: | 60C |
Sample Injection: | 10 uL |
Solvent A: | 100% water |
Solvent B: | 100% acetonitrile |
Analytical Time: | 10 min |
Sample Loop Size: | 15 uL |
Sample Syringe Size: | 100 uL |
Chromatography Type: | HILIC |
Chromatography ID: | CH001748 |
Chromatography Summary: | The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6- port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 uL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 2 min, increased to 0.4 mL/min at 6 min and held for 4 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 80% A, 15% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 2 min, hold for 4 min, resulting in a total analytical run time of 10 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 7.5 min, followed by an equilibration solution of 80% A, 15% B, 5% C for 2.5 min. |
Methods Filename: | 20160120_posC18120kres10mim_ESI_hilicposwash_acore.meth |
Instrument Name: | Dionex UltiMate 3000 |
Column Name: | Thermo Fisher Accucore HILIC (50 x 2.1mm,2.6um) with Thermo accucore HILIC guard cartridge |
Column Temperature: | 60C |
Flow Rate: | 0.35 mL/min for 2 min; linear increase to 0.4 mL/min at 6 min held for 4 min |
Sample Injection: | 10 uL |
Solvent A: | 100% water |
Solvent B: | 100% acetonitrile |
Analytical Time: | 10 min |
Sample Loop Size: | 15 uL |
Sample Syringe Size: | 100 uL |
Chromatography Type: | Reversed phase |