MB Sample ID: SA273658

Local Sample ID:Raw_SmpID_42
Subject ID:SU002821
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004403 AN004404
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Agilent 1260 Agilent 1260
Column Waters Atlantis HILIC (150 x 2.1mm,3um) Waters XBridge Amide (100 x 2.1mm,3.5um)
MS Type ESI ESI
MS instrument type Triple quadrupole Triple quadrupole
MS instrument name ABI Sciex 5500 QTrap ABI Sciex 5500 QTrap
Ion Mode POSITIVE POSITIVE
Units Relative abundance Relative abundance

Chromatography:

Chromatography ID:CH003304
Chromatography Summary:HPLC gradient: The HPLC gradient program begins with an initial condition of 5% solvent A and 95% solvent B, with a flow rate of 0.25 mL/min, which is held for 0.5 minutes to establish system equilibration. The gradient then proceeds as follows: at 6 minutes, the mobile phase composition shifts to 60% solvent A and 40% solvent B for 3minutes; at 10 minutes, it changes back to 5% solvent A and 95% solvent B; at 11 minutes, the flow rate increases to 0.4 mL/min while maintaining a composition of 5% solvent A and 95% solvent B for a duration of 12.5 minutes; and at 24.5 minutes, the flow rate decreases back to 0.25 mL/min. The final condition is maintained for 1 minutes to ensure stability before subsequent analyses.
Instrument Name:Agilent 1260
Column Name:Waters Atlantis HILIC (150 x 2.1mm,3um)
Column Temperature:40°C
Flow Gradient:HPLC gradient: The HPLC gradient program begins with an initial condition of 5% solvent A and 95% solvent B, with a flow rate of 0.25 mL/min, which is held for 0.5 minutes to establish system equilibration. The gradient then proceeds as follows: at 6 minutes, the mobile phase composition shifts to 60% solvent A and 40% solvent B for 3minutes; at 10 minutes, it changes back to 5% solvent A and 95% solvent B; at 11 minutes, the flow rate increases to 0.4 mL/min while maintaining a composition of 5% solvent A and 95% solvent B for a duration of 12.5 minutes; and at 24.5 minutes, the flow rate decreases back to 0.25 mL/min.
Flow Rate:0.250 – 0.400 mL/min
Solvent A:0.1% Formic acid in 10 mM Ammonium Formate (pH ~2.5)
Solvent B:0.1% Formic Acid in Acetonitrile
Chromatography Type:HILIC
  
Chromatography ID:CH003305
Chromatography Summary:HPLC gradient: The HPLC gradient program is initialized with an initial mobile phase composition of 15% solvent A and 85% solvent B at a flow rate of 0.25 mL/min. Over the course of 8 minutes, the mobile phase composition undergoes a transition to 65% solvent A and 35% solvent B. Subsequently, at 8 minutes, the composition shifts to 98% solvent A and 2% solvent B, maintained for 1 minute. The mobile phase reverts back to the initial composition of 15% solvent A and 85% solvent B at 10 minutes. At 12.5 minutes, the flow rate is increased to 0.5 mL/min, while maintaining a constant mobile phase composition of 15% solvent A and 85% solvent B for a period of 2.5 minutes. Finally, at 15 minutes, the flow rate is reduced back to 0.25 mL/min. To ensure system stability, the final condition is maintained for 1 minute prior to subsequent analyses.
Instrument Name:Agilent 1260
Column Name:Waters XBridge Amide (100 x 2.1mm,3.5um)
Column Temperature:40°C
Flow Gradient:HPLC gradient: The HPLC gradient program is initialized with an initial mobile phase composition of 15% solvent A and 85% solvent B at a flow rate of 0.25 mL/min. Over the course of 8 minutes, the mobile phase composition undergoes a transition to 65% solvent A and 35% solvent B. Subsequently, at 8 minutes, the composition shifts to 98% solvent A and 2% solvent B, maintained for 1 minute. The mobile phase reverts back to the initial composition of 15% solvent A and 85% solvent B at 10 minutes. At 12.5 minutes, the flow rate is increased to 0.5 mL/min, while maintaining a constant mobile phase composition of 15% solvent A and 85% solvent B for a period of 2.5 minutes.
Flow Rate:0.250 – 0.400 mL/min
Solvent A:95:5 H2O:Acetonitrile (v:v) with 20mM Ammonium Acetate and 20mM Ammonium Hydroxide (pH 9.0)
Solvent B:Acetonitrile
Chromatography Type:HILIC