Return to study ST002784 main page
MB Sample ID: SA298848
Local Sample ID: | 47_T2 |
Subject ID: | SU002891 |
Subject Type: | Mammal |
Subject Species: | Sus scrofa |
Taxonomy ID: | 9823 |
Age Or Age Range: | ~4 months |
Weight Or Weight Range: | 35-45 kg |
Gender: | Male |
Animal Animal Supplier: | TOPIGS TN70, Tojapigs, the Netherlands |
Animal Light Cycle: | 12 hours |
Animal Water: | ad libitum |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004531 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Agilent 1290 Infinity |
Column | Agilent InfinityLab Poroshell 120 (2.1 x 150 mm, 2.7 um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive Focus |
Ion Mode | NEGATIVE |
Units | Peak area |
Chromatography:
Chromatography ID: | CH003404 |
Chromatography Summary: | 10 µl was loaded onto an Ultra Performance Liquid Chromatograph (UPLC) equipped with a Hydrophilic Interaction Liquid Chromatography (HILIC) column (InfinityLab Poroshell 120 HILIC-Z PEEK-lined 2.1 x 150 mm, 2.7 µm column; Agilent, Santa Clara, USA)) and connected in-line to a Q-exactive Orbitrap Focus (Thermo Fisher Scientific, Massachusetts, USA) mass spectrometer. A linear gradient was built up starting with 90% solvent A (LC-MS grade acetonitrile, Acetonitrile hypergrade for LC-MS LiChrosolv, Supelco (Merck), Germany) and 10% solvent B (10 mM ammonium acetate (LiChropur™, eluent additive for LC-MS, (Merck), Germany), pH 9.3). At 2 min the gradient increased to 60% of solvent B and maintained at 60% until 15 min. The gradient returned to 10% solvent B at 16 min and remained until 25 min. The flow rate was 250 µl/min and the column was kept at 25°C throughout the analysis. The MS operated in negative ion mode, with a spray voltage of 2.9 kV and a temperature of the capillary of 325 °C. Gas settings were as follows: sheath gas 40 and auxiliary gas 15. The vaporizer temperature was set at 300 °C. A full scan (resolution of 70.000 and scan range of m/z 70-1050) was applied. XCalibur version to operate the LC-MS was 4.2.47. |
Instrument Name: | Agilent 1290 Infinity |
Column Name: | Agilent InfinityLab Poroshell 120 (2.1 x 150 mm, 2.7 um) |
Column Temperature: | 25°C |
Flow Gradient: | A linear gradient was built up starting with 90% solvent A and 10% solvent B. At 2 min the gradient increased to 60% of solvent B and maintained at 60% until 15 min. |
Flow Rate: | 250 µl/min |
Solvent A: | 100% acetonitrile |
Solvent B: | 100% water; 10 mM ammonium acetate, pH 9.3 |
Chromatography Type: | HILIC |