Analysis ID | AN001929 | AN001930 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | CE | CE |
Chromatography system | Agilent 1290 Infinity II | Agilent 1290 Infinity II |
Column | a fused silica capillary (50 μm i.d. × 80 cm) | a fused silica capillary (50 μm i.d. × 80 cm) |
MS Type | ESI | ESI |
MS instrument type | TOF | TOF |
MS instrument name | Agilent 6224 TOF | Agilent 6224 TOF |
Ion Mode | POSITIVE | NEGATIVE |
Units | Relative intensity after normalization | Relative intensity after normalization |
MS:
MS ID: | MS001785 |
Analysis ID: | AN001929 |
Instrument Name: | Agilent 6224 TOF |
Instrument Type: | TOF |
MS Type: | ESI |
MS Comments: | nebulizer pressure of 5 psig, dry gas temperature of 300 °C, nitrogen flow of 7 L/min, capillary voltage of 4 kV, fragmentor of 105 V, skimmer of 50 V, Oct RFV of 650 V, acquisition rate of 1.5 spectra/s and mass range of 60-1,000 were utilized. |
Ion Mode: | POSITIVE |
MS ID: | MS001786 |
Analysis ID: | AN001930 |
Instrument Name: | Agilent 6224 TOF |
Instrument Type: | TOF |
MS Type: | ESI |
MS Comments: | capillary voltage of 3.5 kV, fragmentor of 125 V and mass range of 50-1,000 were set in the TOF/MS analysis. During sample analysis, reference masses were applied to ensure the real-time calibration of exact mass measurement. |
Ion Mode: | NEGATIVE |