Return to study ST002075 main page
MB Sample ID: SA196028
Local Sample ID: | 1902 |
Subject ID: | SU002158 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Gender: | Male and female |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003382 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Thermo Vanquish |
Column | Waters Acquity UPLC BEH amide,(150 x 2.1mm,1.7um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE |
Units | peak height |
MS:
MS ID: | MS003149 |
Analysis ID: | AN003382 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | A Thermo Q-Exactive HF Orbitrap MS instrument was operated in positive and negative ESI mdoes respectively with the following parameters: mass range 60−900 m/z; spray voltage 3.6kV (ESI+) and −3kV (ESI−), sheath gas (nitrogen) flow rate 60 units; auxiliary gas (nitrogen) flow rate 25 units, capillary temperature 320 ◦C, full scan MS1 mass resolving power 60,000, data-dependent MSMS (dd-MSMS) 4 scans per cycle, normalized collision energy at 20%, 30%, and 40%, dd-MSMS mass resolving power 15,000. Thermo Xcalibur 4.0.27.19 was used for data acquisition and analysis. Instruments was tuned and calibrated by manufacturer’s recommendations. MSDIAL was used for data processing. Conjugated bile acid peak height was converted to relative concentration in ng/mL by comparing to targeted bile acid dataset ST002073. |
Ion Mode: | POSITIVE |