Return to study ST002411 main page
MB Sample ID: SA241835
Local Sample ID: | 1525 |
Subject ID: | SU002500 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003930 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Vanquish |
Column | Waters Acquity UPLC CSH C18 column (100 mm length × 2.1 mm i.d.; 1.7-µm particle size) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive HF-X Orbitrap |
Ion Mode | NEGATIVE |
Units | peak height |
MS:
MS ID: | MS003668 |
Analysis ID: | AN003930 |
Instrument Name: | Thermo Q Exactive HF-X Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | HESI source conditions are as follows: sheath gas flow 55, auxiliary gas flow 15, sweep gas flow 3, capillary temperature 275°C, S-lens RF level 50, auxiliary gas heater temperature 450 °C, and needle voltage 3500 V and -3500 V for positive and negative ionization mode, respectively. DDA MS/MS spectra were acquired for the top 4 ions. MS scans were collected with 60k resolving power from 120-1700 m/z, AGC target of 106 ions, and maximum accumulation time of 100 ms. MS/MS spectra were collected with 15k resolving power, 1 Da isolation window, normalized collision energy of 20, 30, 60, 2 s dynamic exclusion window, 8×103 AGC target, and 50 ms maximum accumulation time. Spectra were stored in centroid mode. Three rounds of iterative exclusion MS/MS were acquired for each pooled QC sample. |
Ion Mode: | NEGATIVE |