MB Sample ID: SA251302
Local Sample ID: | IKIDS_S_246 |
Subject ID: | SU002599 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004111 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Vanquish |
Column | Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive HF-X Orbitrap |
Ion Mode | POSITIVE |
Units | m/z |
MS:
MS ID: | MS003858 |
Analysis ID: | AN004111 |
Instrument Name: | Thermo Q Exactive HF-X Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Instrument: Thermo Q Exactive HFx Software: Xcalibur 4.1.31.9 for data acquisition; Progenesis QI 2.4 for data preprocessing |
Ion Mode: | POSITIVE |
Capillary Temperature: | 320 °C |
Capillary Voltage: | 3.5 KV |
Collision Energy: | 20-45, ramp |
Collision Gas: | N2 |
Dry Gas Flow: | 55 |
Dry Gas Temp: | 400°C |
Fragmentation Method: | CID |
Ion Spray Voltage: | 3.5kV |
Ionization: | ES+ |
Mass Accuracy: | 5 ppm |
Dataformat: | Profile |
Desolvation Gas Flow: | 55 |
Desolvation Temperature: | 400 oC |