Return to study ST002536 main page
MB Sample ID: SA255000
Local Sample ID: | HC022_N |
Subject ID: | SU002636 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Gender: | Male and female |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004171 | AN004172 | AN004173 | AN004174 |
---|---|---|---|---|
Analysis type | MS | MS | MS | MS |
Chromatography type | HILIC | HILIC | Reversed phase | Reversed phase |
Chromatography system | Shimadzu Nexera X2 | Shimadzu Nexera X2 | Shimadzu Nexera X2 | Shimadzu Nexera X2 |
Column | Waters Atlantis HILIC (150 x 2.1mm,3um,100Å | Phenomenex Luna NH2 (150 x 2 mm,5um;100Å) | Waters ACQUITY UPLC BEH C8 Column (2.1 mm X 100 mm,1.7 µm,130Å) | Waters ACQUITY UPLC BEH C18 Column (2.1 mm X 150 mm,1.7 µm,130Å) |
MS Type | ESI | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Plus Orbitrap | Thermo Exactive Plus Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE | POSITIVE | NEGATIVE |
Units | peak area | peak area | peak area | peak area |
MS:
MS ID: | MS003918 |
Analysis ID: | AN004171 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | MS analyses were carried out using electrospray ionization in the positive ion mode using full scan analysis over 70-800 m/z at 70,000 resolution and 3 Hz data acquisition rate. Other MS settings were: sheath gas 40, sweep gas 2, spray voltage 3.5 kV, capillary temperature 350°C, S-lens RF 40, heater temperature 300°C, microscans 1, automatic gain control target 1e6, and maximum ion time 250 ms. Raw data were processed using TraceFinder 3.3 software (Thermo Fisher Scientific; Waltham, MA) and Progenesis QI (Nonlinear Dynamics; Newcastle upon Tyne, UK). Metabolite identities were confirmed using authentic reference standards. |
Ion Mode: | POSITIVE |
MS ID: | MS003919 |
Analysis ID: | AN004172 |
Instrument Name: | Thermo Q Exactive Plus Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | MS analyses were carried out using electrospray ionization in the negative ion mode using full scan analysis over m/z 70-750 at 70,000 resolution and 3 Hz data acquisition rate. Additional MS settings were: ion spray voltage, -3.0 kV; capillary temperature, 350°C; probe heater temperature, 325 °C; sheath gas, 55; auxiliary gas, 10; and S-lens RF level 50. Raw data were processed using TraceFinder 3.3 software (Thermo Fisher Scientific; Waltham, MA) and Progenesis QI (Nonlinear Dynamics; Newcastle upon Tyne, UK). Metabolite identities were confirmed using authentic reference standards. |
Ion Mode: | NEGATIVE |
MS ID: | MS003920 |
Analysis ID: | AN004173 |
Instrument Name: | Thermo Exactive Plus Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | MS analyses were carried out using electrospray ionization in the positive ion mode using full scan analysis over 200–1100 m/z at 70,000 resolution and 3 Hz data acquisition rate. Other MS settings were: sheath gas 50, in source CID 5 eV, sweep gas 5, spray voltage 3 kV, capillary temperature 300°C, S-lens RF 60, heater temperature 300°C, microscans 1, automatic gain control target 1e6, and maximum ion time 100 ms. Representative standards from each lipid class were used to identify lipids based on RT and m/z patterns, and characteristic product ion spectra. Lipid identities were denoted by total acyl carbon number and total number of double bond number. |
Ion Mode: | POSITIVE |
MS ID: | MS003921 |
Analysis ID: | AN004174 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | MS analyses were carried out using electrospray ionization in the negative ion mode using full scan analysis over m/z 70-850 at 70,000 resolution and 3 Hz data acquisition rate. Additional MS settings were: ion spray voltage, -3.5 kV; capillary temperature, 320°C; probe heater temperature, 300 °C; sheath gas, 45; auxiliary gas, 10; and S-lens RF level 60. Raw data were processed using TraceFinder 3.3 software (Thermo Fisher Scientific; Waltham, MA) and Progenesis QI (Nonlinear Dynamics; Newcastle upon Tyne, UK). Metabolite identities were confirmed using authentic reference standards. |
Ion Mode: | NEGATIVE |