Combined analysis:
Analysis ID | AN004390 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Vanquish™ Flex UHPLC |
Column | Merck SeQuant ZIC-pHILIC (150 x 2.1mm,5um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap |
Ion Mode | UNSPECIFIED |
Units | a.u. |
MS:
MS ID: | MS004139 |
Analysis ID: | AN004390 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | MS data acquisition was performed using a QExactive benchtop orbitrap mass spectrometer equipped with an Ion Max source and a HESI II probe (Thermo Fisher Scientific, San Jose, CA) and was performed in positive and negative ionization mode in a range of m/z = 70–1000, with the resolution set at 70,000, the AGC target at 1x106, and the maximum injection time (Max IT) at 20 msec. A narrower scan in positive mode at m/z = 600-800 was used for more specific detection of thyroxine hormones, the resolution was set at 70,000, the AGC target was 5x105, and the max IT was 100 msec. For polar metabolites HESI conditions were: Sheath gas frow rate: 35; Aug gas flow rate: 8; Sweet gas flow rate: 1; Spray voltage: 3.5kV (pos), 2.8kV (neg); Capillary temperature: 320ºC; S-lens RF: 50; Aux gas heater temperature: 350 ºC. |
Ion Mode: | UNSPECIFIED |