MB Sample ID: SA289284
Local Sample ID: | Inf5_infected_NA_5__015 |
Subject ID: | SU002854 |
Subject Type: | Cultured cells |
Subject Species: | Rickettsia parkeri; Homo sapiens |
Taxonomy ID: | 35792; 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004454 | AN004455 | AN004456 | AN004457 |
---|---|---|---|---|
Analysis type | MS | MS | MS | MS |
Chromatography type | HILIC | HILIC | Reversed phase | Reversed phase |
Chromatography system | Thermo Vanquish | Thermo Vanquish | Thermo Vanquish | Thermo Vanquish |
Column | Waters ACQUITY UPLC BEH Amide (150 x 2.1mm,1.7um) | Waters ACQUITY UPLC BEH Amide (150 x 2.1mm,1.7um) | Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um) | Waters ACQUITY UPLC CSH C18 (100 x 2.1mm,1.7um) |
MS Type | ESI | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE | NEGATIVE | POSITIVE | NEGATIVE |
Units | counts, height | counts, height | counts, height | counts, height |
MS:
MS ID: | MS004201 |
Analysis ID: | AN004454 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Spectra were collected using a Thermo Q Exactive HF Hybrid Quadrupole-Orbitrap mass spectrometer in both positive and negative mode ionization (separate injections). Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from the in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.0 Da. Collision energy was NCE 20, 50, 100 eV. |
Ion Mode: | POSITIVE |
MS ID: | MS004202 |
Analysis ID: | AN004455 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Spectra were collected using a Thermo Q Exactive HF Hybrid Quadrupole-Orbitrap mass spectrometer in both positive and negative mode ionization (separate injections). Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from the in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.0 Da. Collision energy was NCE 20, 50, 100 eV. |
Ion Mode: | NEGATIVE |
MS ID: | MS004203 |
Analysis ID: | AN004456 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Spectra were collected using a Thermo Q Exactive HF Hybrid Quadrupole-Orbitrap mass spectrometer in both positive and negative mode ionization (separate injections). Full MS-ddMS2 data was collected. Mass range was 220-1600 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.0 Da. Collision energy was NCE 20, 40, 60 eV. |
Ion Mode: | POSITIVE |
MS ID: | MS004204 |
Analysis ID: | AN004457 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Spectra were collected using a Thermo Q Exactive HF Hybrid Quadrupole-Orbitrap mass spectrometer in both positive and negative mode ionization (separate injections). Full MS-ddMS2 data was collected. Mass range was 220-1600 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.0 Da. Collision energy was NCE 20, 40, 60 eV. |
Ion Mode: | NEGATIVE |