Return to study ST003006 main page
MB Sample ID: SA326864
Local Sample ID: | BAT_6C_chow_neg_9 |
Subject ID: | SU003120 |
Subject Type: | Mammal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004937 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Vanquish Horizon |
Column | Waters ACQUITY UPLC BEH Amide (100 x 2.1mm,1.7um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo orbitrap exploris 240 |
Ion Mode | NEGATIVE |
Units | Peak area |
MS:
MS ID: | MS004680 |
Analysis ID: | AN004937 |
Instrument Name: | Thermo orbitrap exploris 240 |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | ESI source parameters were set as follows: spray voltage, 3500 V or −2800 V, in positive or negative modes, respectively; vaporizer temperature, 350 °C; sheath gas, 50 arb; aux gas, 10 arb; ion transfer tube temperature, 325 °C. The full scan was set as: orbitrap resolution, 60,000; maximum injection time, 100 ms; scan range, 70–1050 Da. The ddMS2 scan was set as: orbitrap resolution, 30,000; maximum injection time, 60 ms; top N setting, 6; isolation width, 1.0 m/z; HCD collision energy (%), 30; Dynamic exclusion mode was set as auto. The metabolites was quantified by Compound Discoverer 3.3. |
Ion Mode: | NEGATIVE |