Summary of Study ST000091

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000083. The data can be accessed directly via it's Project DOI: 10.21228/M8T884 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files
Study IDST000091
Study TitleQuantitative Metabolomics by 1H-NMR and LC-MS/MS Confirms Altered Metabolic Pathways in Diabetes
Study TypePre- and Post- insulin study with matched controls
Study SummaryWe obtained plasma samples from 7 c-peptide negative type 1 diabetic individuals (T1D) and 7 non-diabetic controls (Con) that were matched for age (T1D = 31.1±2.9 yrs, Con = 30.2±3.4 yrs), body mass (T1D = 80.2±4.7kg, Con = 81.9±7.4 kg) and BMI (T1D = 26.5±1.2 kg/m2, Con = 25.2±1.3 kg/m2). Type 1 diabetic people were studied while treated with insulin and also after 8 hours of insulin deprivation
Institute
Mayo Clinic
DepartmentEndocrinology
Last NameNair
First NameSreekumaran
Address--
Emailnair@mayo.edu
Phone--
Submit Date2014-07-15
Num Groups2
Raw Data AvailableNo
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2014-07-26
Release Version1
Sreekumaran Nair Sreekumaran Nair
https://dx.doi.org/10.21228/M8T884
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000145
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity
Column Waters Acquity BEH C18 (150 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name Thermo Quantum Ultra
Ion Mode POSITIVE
Units uM

Chromatography:

Chromatography ID:CH000103
Chromatography Summary:High resolution separation was done using an Acquity UPLC system, injecting 1 l of derviatized solution, with a UPLC BEH C18 1.7 micron 2.1150 mm column from Waters. Column flow was set to 400 l/min with a gradient from 99.9%A to 98%B where buffer A is 1% acetonitrile in 0.1% formic acid and buffer B is 100% acetonitrile. A column temp of 43 degrees Celsius and a sample tray temp of 6 degree Celsius.
Instrument Name:Waters Acquity
Column Name:Waters Acquity BEH C18 (150 x 2.1mm,1.7um)
Column Temperature:43
Flow Gradient:99.9% A to 98% B
Flow Rate:401 µl/min
Solvent A:99% water/1% acetonitrile; 0.1% formic acid
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase