Summary of Study ST000140

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000125. The data can be accessed directly via it's Project DOI: 10.21228/M8DW2B This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000140
Study TitleCell Rinsing Solution Comparison
Study TypeCell Rinsing Solution Comparison
Study SummaryAliquots of Jurkat T-lymphocyte cells were washed with 5 different rinsing solutions (0.3% amm. formate, 0.3% amm. acetate, 0.9% NaCl, 1 M PBS, 100 mM PBS) and ran in triplicate to monitor the effect of ion suppression on the electrospray ionization signal.
Institute
University of Florida
DepartmentDept. of Chemistry/SECIM
LaboratoryBiomedical Mass Spectrometry Lab
Last NameUlmer
First NameCandice
AddressP.O. Box 117200, Gainesville, FL 32611
Emailczulmer@chem.ufl.edu
Phone(352) 392-0515
Submit Date2015-03-02
Num Groups5
Total Subjects15
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Uploaded File Size3.1 G
Analysis Type DetailLC-MS
Release Date2016-03-03
Release Version1
Candice Ulmer Candice Ulmer
https://dx.doi.org/10.21228/M8DW2B
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000222
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS
Column ACE Excel 2 C18-PFP (100 x 2.1mm, 2um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units Peak area

MS:

MS ID:MS000173
Analysis ID:AN000222
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
Ion Mode:POSITIVE
Capillary Temperature:320
Dry Gas Flow:45
Mass Accuracy:<2 ppm
Spray Voltage:3.5 kV
Dataformat:.raw
Scan Range Moverz:70-700
Skimmer Voltage:15 V
Acquisition Parameters File:QE1_CZU_17_SEQUENCE