Summary of Study ST002427

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001561. The data can be accessed directly via it's Project DOI: 10.21228/M8XD8F This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002427
Study TitleDansylated Human Plasma at Known Light-Heavy Mixing Ratios
Study TypeMS isotopic labeling analysis
Study SummaryPooled human plasma was split and tagged with either light or heavy dansyl chloride and then recombined at known light-to-heavy ratios for LC-MS analysis and subsequent testing of isotopic labeling software.
Institute
University of North Carolina Greensboro
Last NameZhang
First NameQibin
Address600 Laureate Way, Suite 2203, Kannapolis, NC 28081
Emailq_zhang2@uncg.edu
Phone+1 (704) 250-5810
Submit Date2022-12-21
Raw Data AvailableYes
Raw Data File Type(s)d
Analysis Type DetailLC-MS
Release Date2023-01-20
Release Version1
Qibin Zhang Qibin Zhang
https://dx.doi.org/10.21228/M8XD8F
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003950
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units unitless L/H ratio

MS:

MS ID:MS003685
Analysis ID:AN003950
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:A Thermo Scientific Vanquish Horizon UHPLC System coupled with a Thermo Scientific Q Exactive HF Orbitrap was used to conduct LC-(+)FTMS analysis. Ionization for FTMS analysis utilized a HESI source in the positive mode with spray voltage 3 kV, capillary temperature 300 ºC, sheath gas flowrate 35 au, aux gas flowrate 10 au, spare gas flowrate 2 au, max spray current 100 V, probe heater temperature 300 ºC, and S-lens RF level 50 V. FTMS scanning was full MS1 in the range 250 to 1000 m/z with maximum IT 200 ms, AGC target 5x105, and mass resolution 60,000. MS-DIAL v4.9.2 was used to perform peak alignment and annotation. The novel MS-DIAL addon Peak Pair Pruner was used to perform peak pair validation and quantitation with light-to-heavy ratio reporting and corrections for background peaks and isotopic overlap between the light and heavy tags.
Ion Mode:POSITIVE
Capillary Temperature:300 C
Ion Spray Voltage:3000 V
Analysis Protocol File:DansylatedHumanPlasma.pdf