Summary of Study ST000785

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000571. The data can be accessed directly via it's Project DOI: 10.21228/M8VM2M This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000785
Study TitlePharmacometabolomics of L-Carnitine Treatment Response Phenotypes in Patients with Septic Shock
Study Typemultiple timepoints; patients with severe sepsis or septic shock
Study Summaryphase I study of L-carnitine infusion for the treatment of vasopressor-dependent shock
Institute
University of Michigan
DepartmentClinical Pharmacy
LaboratoryThe NMR Metabolomics Laboratory (Stringer)
Last NameStringer
First NameKathleen
AddressUniversity Michigan, 2900 Huron Parkway, Ann Arbor, MI 48105
EmailNMRmetabolomics@umich.edu
Phonenone
Submit Date2016-12-08
Num Groups2
Total Subjects31
Raw Data AvailableYes
Raw Data File Type(s)fid
Analysis Type DetailNMR
Release Date2017-10-03
Release Version1
Kathleen Stringer Kathleen Stringer
https://dx.doi.org/10.21228/M8VM2M
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


NMR:

NMR ID:NM000105
Analysis ID:AN001244
Instrument Name:Agilent 500/54 Premium Shielded VNMRS system
Instrument Type:FT-NMR
NMR Experiment Type:1D 1H
NMR Comments:Experiment dates: 11/7/2013, 11/11/2013, 11/19/2013
Standard Concentration:10% (~0.5mM)
Spectrometer Frequency:500 MHz
NMR Probe:ONE-Probe
NMR Solvent:D2O
NMR Tube Size:5mm
Shimming Method:Auto shim (gradient shimming)
Pulse Sequence:1D-NOESY
Water Suppression:saturation at 80 Hz induced field strength
Pulse Width:5.5ms
Power Level:around -178Hz
Chemical Shift Ref Cpd:Formate
Temperature:25°C
Number Of Scans:32
Baseline Correction Method:manual
Chemical Shift Ref Std:Calcium Formate