Summary of Study ST000892

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000621. The data can be accessed directly via it's Project DOI: 10.21228/M8D97C This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000892
Study TitleComparison of T2 filters for quantification of small molecules in plasma by 1D NMR
Study TypeNMR development for metabolomics
Study SummaryA pooled plasma sample obtained from Red Cross has been seperated into 20 samples. 10 of them were simply mixed with a D2O phosphate buffer, while the other 10 were centrifuged first, then dried and resuspended in a 90/10 H2O/D2O phosphate buffer. The 20 NMR samples were analysed by NMR (4 spectra per sample, but different pulse sequences and parameters were used for each group). Standard addition was used in order to measure the concentration of 5 metabolites: valine, lactate, glucose, phenylalanine and formate.
Institute
University of Georgia
DepartmentComplex Carbohydrate Research Center
Last NameEdison
First NameArthur
Address315 Riverbend Road 30602 Athens GA
Emailaedison@uga.edu
Phone+1-706-542-8156
Submit Date2017-09-01
Num Groups2
Total Subjects20
Raw Data AvailableYes
Raw Data File Type(s)fid
Analysis Type DetailNMR
Release Date2017-11-15
Release Version1
Arthur Edison Arthur Edison
https://dx.doi.org/10.21228/M8D97C
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


NMR:

NMR ID:NM000110
Analysis ID:AN001454
Instrument Name:FT NMR
Instrument Type:FT-NMR
NMR Experiment Type:1D 1H
NMR Comments:For intact plasma | For intact plasma | For intact plasma | For intact plasma | For extracted plasma | For extracted plasma | For extracted plasma | For extracted plasma
Field Frequency Lock:2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz
Spectrometer Frequency:600 MHz
NMR Solvent:H2O+D2O (90/10) phosphate buffer | H2O+D2O (90/10) phosphate buffer | H2O+D2O (90/10) phosphate buffer | H2O+D2O (90/10) phosphate buffer | D2O phosphate buffer | D2O phosphate buffer | D2O phosphate buffer | D2O phosphate buffer
NMR Tube Size:5 mm
Shimming Method:Topshim | Topshim | Topshim | Topshim | Topshim | Topshim | Topshim | Topshim Pulse Sequence | cpmgpr1d | cpmgpr1d | cpmgpr1d | PROJECTpr1d-1 | noesypr1d | zgpurge-gprefocus-2 | cpmgpr1d | PROJECTpr1d-1
Water Suppression:Presaturation
Pulse Width:19.94 μs | 19.94 μs | 19.94 μs | 19.94 μs | 11.82 μs | 11.82 μs | 11.82 μs | 11.82 μs
Power Level:5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W
Receiver Gain:90.5 | 90.5 | 90.5 | 90.5 | 90.5 | 90.5 | 90.5 | 90.5
Offset Frequency:2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz
Presaturation Power Level:4.13E-05 | 4.13E-05 | 4.13E-05 | 4.13E-05 | 4.13E-05 | 4.13E-05 | 4.13E-05 | 4.13E-05
Chemical Shift Ref Cpd:DSS | DSS | DSS | DSS | DSS | DSS | DSS | DSS
Temperature:310 K | 311 K | 312 K | 313 K | 314 K | 315 K | 316 K | 317 K
Number Of Scans:32 | 32 | 32 | 32 | 32 | 32 | 32 | 32
Dummy Scans:4 | 4 | 4 | 4 | 4 | 4 | 4 | 4
Acquisition Time:3.067 | 3.067 | 3.067 | 3.067 | 3.067 | 3.067 | 3.067 | 3.067
Relaxation Delay:4 s | 4 s | 4 s | 4 s | 5 s | 6 s | 7 s | 8 s
Spectral Width:20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm
Num Data Points Acquired:73728 | 73728 | 73728 | 73728 | 73728 | 73728 | 73728 | 73728
Real Data Points:36864 | 36864 | 36864 | 36864 | 36864 | 36864 | 36864 | 36864
Line Broadening:0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz
Zero Filling:65536 | 65536 | 65536 | 65536 | 65536 | 65536 | 65536 | 65536
Apodization:Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian | Gaussian
Baseline Correction Method:Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2
Chemical Shift Ref Std:DSS | DSS | DSS | DSS | DSS | DSS | DSS | DSS