Summary of Study ST002558

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001650. The data can be accessed directly via it's Project DOI: 10.21228/M8DX4F This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002558
Study TitleExtraction and Untargeted Analysis of Metabolome from Undemineralised Cortical Bone Matrix for Forensic Application
Study SummaryUntargeted metabolomics has become the gold standard for the profiling of low-molecular-weight compounds. Recently, metabolomics has shown great potential in forensic science in the field of toxicology and postmortem interval estimation. The current study aims to evaluate three extraction protocol and four liquid chromatography coupled with mass spectrometry assays that could offer a valuable tool to identify biomarkers for PMI estimation. One fragment for anterior human skeletal tibia from a 82 years old male individual belonging to the Forensic Anthropology Center - Texas State University collection was powdered and extracted in five replicates to be extracted according to a the biphasic chloroform/methanol/water protocol and two single phase protocols based on methanol/water and methanol/acetonitrile/water. Formal analysis was carried out ThermoFisher Ultimate 3000 HPLC in hydrophilic interaction (HILIC) and reverse phase (RP) liquid chromatography coupled with SCIEX 6600 TripleTOF Q-TOF mass spectrometer.
Institute
University of Central Lancashire
DepartmentNatural Sciences
Last NameBonicelli
First NameAndrea
Address1 North Cliff street
Emailabonicelli@uclan.ac.uk
Phone07383974949
Submit Date2023-03-15
Num Groups3
Total Subjects1
Num Males1
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2024-03-17
Release Version1
Andrea Bonicelli Andrea Bonicelli
https://dx.doi.org/10.21228/M8DX4F
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004214 AN004215 AN004216 AN004217
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000
Column Agilent InfinityLab Poroshell 120 HILIC-Z (150 x 2.1mm,2.7um) Agilent InfinityLab Poroshell 120 HILIC-Z (150 x 2.1mm,2.7um) Thermo Accucore C18 (150 x 2.1mm,2.6um) Thermo Accucore C18 (150 x 2.1mm,2.6um)
MS Type ESI ESI ESI ESI
MS instrument type QTOF QTOF QTOF QTOF
MS instrument name ABI Sciex 6600 TripleTOF ABI Sciex 6600 TripleTOF ABI Sciex 6600 TripleTOF ABI Sciex 6600 TripleTOF
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units area integration area integration area integration area integration

Chromatography:

Chromatography ID:CH003124
Chromatography Summary:Hilic ESI+
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Agilent InfinityLab Poroshell 120 HILIC-Z (150 x 2.1mm,2.7um)
Column Temperature:50 °C
Flow Gradient:0 min 98% B 3 min 98% B 23 min 5% B 24 min 5% B 24 min 98% B
Flow Rate:0.25 ml/min
Sample Injection:5 μL
Solvent A:10 mM ammonium formate and 0.1 % formic acid in water
Solvent B:9:1 acetonitrile / 10 mM ammonium formate and 0.1 % formic acid in water
Chromatography Type:HILIC
  
Chromatography ID:CH003125
Chromatography Summary:Hilic ESI-
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Agilent InfinityLab Poroshell 120 HILIC-Z (150 x 2.1mm,2.7um)
Column Temperature:50 °C
Flow Gradient:0 min 96% B 2 min 96% B 22 min 65% B 24 min 65% B 24 min 96% B
Flow Rate:0.25 ml/min
Sample Injection:5 μL
Solvent A:10 mM ammonium acetate, pH 9 with ammonium hydroxide and 10 µM medronic acid
Solvent B:85:15 acetonitrile / 10 mM ammonium acetate, pH 9 with ammonium hydroxide and 10 µM medronic acid in water
Chromatography Type:HILIC
  
Chromatography ID:CH003126
Chromatography Summary:RP ESI+
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Thermo Accucore C18 (150 x 2.1mm,2.6um)
Column Temperature:40 °C
Flow Gradient:0 min 5% B 1 min 5% B 8 min 100% B 10 min 100% B 10 min 5% B
Flow Rate:0.3 ml/min
Sample Injection:5 μL
Solvent A:0.1 % formic acid in water
Solvent B:0.1 % formic acid in 98:2 acetonitrile / water
Chromatography Type:Reversed phase
  
Chromatography ID:CH003127
Chromatography Summary:RP ESI-
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Thermo Accucore C18 (150 x 2.1mm,2.6um)
Column Temperature:40 °C
Flow Gradient:0 min 5% B 1 min 5% B 8 min 100% B 10 min 100% B 10 min 5% B
Flow Rate:0.3 ml/min
Sample Injection:5 μL
Solvent A:0.1 % formic acid in water
Solvent B:0.1 % formic acid in 98:2 acetonitrile / water
Chromatography Type:Reversed phase
  logo