Return to study ST002439 main page

MB Sample ID: SA243715

Local Sample ID:PZFe_HHEAR_6
Subject ID:SU002528
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002528
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
PZFe_HHEAR_6SA243715FL030593HHEAR PoolSample Type

Collection:

Collection ID:CO002521
Collection Summary:NA
Sample Type:Blood (serum)
Storage Conditions:-80℃

Treatment:

Treatment ID:TR002540
Treatment Summary:NA

Sample Preparation:

Sampleprep ID:SP002534
Sampleprep Summary:Serum samples (600 µL per tube) were shipped from Kaiser Permanente Mosswood Research Lab to the NC HHEAR Hub on dry ice. The NC HHEAR Hub thawed and transferred 50 µL of the study samples to a new set of tubes and used them for the analysis. An additional 10 µL was taken from the original study sample and transferred to another tube to make the total study pool for this project, and then distributed with 50 µL per aliquot, used as Quality control study pools (QC study pools) throughout the whole analysis. All sample aliquots (50 µL each) and QC study pools (50 µL each) were stored at -80° C until the day for sample preparation. HHEAR reference plasma (50 µL each) and NIST serum (909c) reference material (50 µL each) were provided by the NC HHEAR Hub. The LC-MS grade water (50 µL) was used as blank. Sample preparation was conducted by batch, and all samples (except blank) were thawed at 4°C overnight before the preparation. Samples, including study samples, study pool samples, HHEAR reference plasma, NIST reference serum, and blanks, were mixed with 400 µL methanol containing 500 ng/mL tryptophan-d5 as internal standard and vortexed by a multiple tube vortex mixer for 2 min at 5000 rpm at room temperature. All samples were centrifuged at 16,000 rcf for 10 min at 4°C. The supernatant (350 µL) was transferred into a pre-labeled 2.0 mL Lo-bind Eppendorf tube, dried by a SpeedVac overnight, and stored at -80° C. For immediate analysis, 100 µL of water-methanol solution (95:5, v/v) was used to reconstitute the dried extracts. Samples were thoroughly mixed on a multiple tube vortex mixer for 10 min at 5000 rpm at room temperature and then centrifuged at 4°C for 10 min at 16,000 rcf. The supernatant was transferred to pre-labeled autosampler vials for data acquisition by LC-MS.
Processing Storage Conditions:4℃
Extraction Method:Vortex with methanol containing 500ng/ml tryptophan-d5 as internal standard
Extract Storage:-80℃
Sample Resuspension:Water-Methanol (95:5, v/v)
Sample Spiking:Tryptophan-d5 stock solution at 500 ng/mL

Combined analysis:

Analysis ID AN003976
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap
Ion Mode POSITIVE
Units m/z

Chromatography:

Chromatography ID:CH002939
Chromatography Summary:Reverse phase
Instrument Name:Thermo Vanquish
Column Name:Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
Column Pressure:6000-10000
Column Temperature:50 oC
Flow Gradient:Time(min) Flow Rate %A %B Curve 1. 0 0.4 99.0 1.0 5 2. 1.00 0.4 99.0 1.0 5 3. 16.00 0.4 1.0 99.0 5 4. 19.00 0.4 1.0 99.0 5 5. 19.50 0.4 99.0 1.0 5 6. 22.00 0.4 99.0 1.0 5
Flow Rate:0.4 mL/min
Injection Temperature:8oC
Solvent A:Water with 0.1% Formic Acid
Solvent B:Methanol with 0.1% Formic Acid
Chromatography Type:Reversed phase

MS:

MS ID:MS003710
Analysis ID:AN003976
Instrument Name:Thermo Q Exactive HF-X Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:Instrument: Thermo Q Exactive HFx Software: Xcalibur 4.1.31.9 for data acquisition; Progenesis QI 2.4 for data preprocessing
Ion Mode:POSITIVE
Capillary Temperature:320 °C
Capillary Voltage:3.5 KV
Collision Energy:20-45, ramp
Collision Gas:N2
Dry Gas Flow:55
Dry Gas Temp:400°C
Fragment Voltage:ES+
Fragmentation Method:CID
Ion Spray Voltage:3.5kV
Ionization:ES+
Mass Accuracy:5 ppm
Dataformat:Profile
Desolvation Gas Flow:55
Desolvation Temperature:400 oC