MB Sample ID: SA297371

Local Sample ID:2021-02-01 JL14 SO WE 2 MRM LM TR3_1-E
Subject ID:SU002885
Subject Type:Amphibian
Subject Species:Xenopus laevis
Taxonomy ID:8355

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002885
Subject Type:Amphibian
Subject Species:Xenopus laevis
Taxonomy ID:8355

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
2021-02-01 JL14 SO WE 2 MRM LM TR3_1-ESA297371FL036050RESample type

Collection:

Collection ID:CO002878
Collection Summary:The Spemann-Mangold Organizer (SMO) was lineage-traced under epifluorescence and isolated in a 2% (w/v) agarose-coated Petri dish containing 50% SS. The dissected SMO tissues and the remainder embryo (RE) were collected in LoBind Eppendorf tubes separately, and processed using our established protocols for metabolomic analysis. For MS-based metabolomics of the TCA cycle and glycolysis, 10 dissections of SMO and RE were pooled as 1 BR. For targeted quantification of reduced and oxidized glutathione, 20 tissues of SMO and RE were dissected and pooled as one BR. The dissected tissues were stored in LC-MS-grade methanol at −80 °C until sample processing.
Sample Type:Dissected organizer from Xenopus laevis

Treatment:

Treatment ID:TR002894
Treatment Summary:NA

Sample Preparation:

Sampleprep ID:SP002891
Sampleprep Summary:The supernatant was transferred to a new 1.5 mL Eppendorf vial, vacuum-dried at 20 °C, and reconstituted in 45 uL of LC-MS grade water. For GSH/GSSG analysis, the supernatant was reconstituted in 45 uL of 50% ACN. The supernatant was centrifuged at 13,000 g for 10 min at 4 °C, then stored at -80 °C until LC-MS analysis.
Sampleprep Protocol Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx

Combined analysis:

Analysis ID AN004522 AN004523 AN004524
Analysis type MS MS MS
Chromatography type Ion exchange Ion exchange HILIC
Chromatography system Waters ACQUITY I-Class Waters Acquity I-Class Waters Acquity I-Class
Column Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um) Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um) ACQUITY UPLC BEH amide (100 x 1mm, 1.7um)
MS Type ESI ESI ESI
MS instrument type QTOF QTOF QTOF
MS instrument name Bruker timsTOF PRO Bruker timsTOF PRO Bruker timsTOF PRO
Ion Mode NEGATIVE NEGATIVE POSITIVE
Units Counts Counts Counts

Chromatography:

Chromatography ID:CH003396
Chromatography Summary:A 2.5 µL of the metabolite extract was loaded onto a reversed-phase LC column featuring anion-exchange chemistry (Atlantis PREMIER BEH C18 AX Column, Waters, 1.7 µm, 2.1 × 100 mm) and separated at 500 uL/min in a micro-flow LC system (ACQUITY I-Class, Waters) using a gradient of buffer A (100% ACN) and buffer B (95% water, 5% ACN with 1 mM ammonium acetate; pH 7.8) as follows: 5% was held 0–¬0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min. To recover and recondition the analytical column before the next sample injection, solvent B was then decreased to 5% over 2 min and the column was rinsed for 7 min. The column temperature was set to 30 °C.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters ACQUITY I-Class
Column Name:Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um)
Column Temperature:30
Flow Gradient:5% B was held 0–0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min
Flow Rate:500 uL/min
Solvent A:100% ACN
Solvent B:95% water/5% acetonitrile; 1 mM ammonium acetate; pH 7.8
Chromatography Type:Ion exchange
  
Chromatography ID:CH003397
Chromatography Summary:A 2.5 µL of the metabolite extract was loaded onto a reversed-phase LC column featuring anion-exchange chemistry (Atlantis PREMIER BEH C18 AX Column, Waters, 1.7 µm, 2.1 × 100 mm) and separated at 500 uL/min in a micro-flow LC system (ACQUITY I-Class, Waters) using a gradient of buffer A (100% ACN) and buffer B (95% water, 5% ACN with 1 mM ammonium acetate; pH 7.8) as follows: 5% was held 0–¬0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min. To recover and recondition the analytical column before the next sample injection, solvent B was then decreased to 5% over 2 min and the column was rinsed for 7 min. The column temperature was set to 30 °C.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters Acquity I-Class
Column Name:Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um)
Column Temperature:30
Flow Gradient:5% B was held 0–0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min
Flow Rate:500 uL/min
Solvent A:100% ACN
Solvent B:95% water/5% acetonitrile; 1 mM ammonium acetate; pH 7.8
Chromatography Type:Ion exchange
  
Chromatography ID:CH003398
Chromatography Summary:A 2-µL volume of the metabolite extract was loaded onto an ACQUITY UPLC BEH amide column (130 A, 1.7 µm, 1 mm × 100 mm) and separated at 45 ℃ using a mobile phase delivered at 130 uL/min in a micro-LC system (Waters ACQUITY I-Class). The separation was performed in buffer A (water containing 0.1% v/v FA) with a gradient of buffer B (100% ACN with 0.1% v/v FA) as follows: 1% A for 0.05 min, then ramped to 80% A over 3.20 min, held at 80% A for 2 min, ramped to 1% A in 1.75 min. To recover and condition the analytical column for the next sample injection, buffer A was held at 1% for 7 min.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters Acquity I-Class
Column Name:ACQUITY UPLC BEH amide (100 x 1mm, 1.7um)
Column Temperature:45
Flow Gradient:1% A for 0.05 min, then ramped to 80% A over 3.20 min, held at 80% A for 2 min, ramped to 1% A in 1.75 min
Flow Rate:130 uL/min
Solvent A:water containing 0.1% v/v FA
Solvent B:100% acetonitrile; 0.1% v/v FA
Chromatography Type:HILIC

MS:

MS ID:MS004269
Analysis ID:AN004522
Instrument Name:Bruker timsTOF PRO
Instrument Type:QTOF
MS Type:ESI
MS Comments:The metabolites were ionized in a micro-flow electrospray source equipped with an ionBooster-ESI device with the following settings: end plate, 400 V; capillary voltage, 1,000 V; charge voltage, 300 V; vaporizer temperature, 240 °C; sheath gas, 1.5 L/min; nebulizer, 4.1 bar; dry gas, 2.6 L/min; dry temperature, 200 °C. The ions were detected in the negative ionization mode using the following settings: survey (MS1) low- and middle-mass range, m/z 20–300 and m/z 50–1,300, respectively; spectral acquisition rate, 2 Hz. A list of metabolite ions was targeted using multiple reaction monitoring. Ions were isolated with 4-Da window (± 2 Da) for collision-induced dissociation (CID). The low-mass range (m/z 20–300) employed the following optimization for collision energy and m/z tuning: 12 eV at 115.0026; 12 eV at 145.0132; 12 eV at 168.9897; 12 eV at 184.9846; 10 eV at 191.0186.
Ion Mode:NEGATIVE
Analysis Protocol File:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
  
MS ID:MS004270
Analysis ID:AN004523
Instrument Name:Bruker timsTOF PRO
Instrument Type:QTOF
MS Type:ESI
MS Comments:The metabolites were ionized in a micro-flow electrospray source equipped with an ionBooster-ESI device with the following settings: end plate, 400 V; capillary voltage, 1,000 V; charge voltage, 300 V; vaporizer temperature, 240 °C; sheath gas, 1.5 L/min; nebulizer, 4.1 bar; dry gas, 2.6 L/min; dry temperature, 200 °C. The ions were detected in the negative ionization mode using the following settings: survey (MS1) low- and middle-mass range, m/z 20–300 and m/z 50–1,300, respectively; spectral acquisition rate, 2 Hz. A list of metabolite ions was targeted using multiple reaction monitoring. Ions were isolated with 4-Da window (± 2 Da) for collision-induced dissociation (CID). The middle-mass range (m/z 50–1,300) employed the following optimization for collision energy and m/z tuning: 15 eV at 338.9877; 18 eV at 346.0553; 25 eV at 426.0216; 20 eV at 505.9879.
Ion Mode:NEGATIVE
Analysis Protocol File:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
  
MS ID:MS004271
Analysis ID:AN004524
Instrument Name:Bruker timsTOF PRO
Instrument Type:QTOF
MS Type:ESI
MS Comments:The GSH and GSSG ion signals were measured in the positive ion mode and detected on the timsTOF PRO (Bruker) mass spectrometer with the following parameters: mass range, m/z 100–800; spectral acquisition rate, 4 Hz; scan mode, MRM (GSH: m/z 308.0911, charge state +1, retention time 2.5 min, width 4 Da (+/– 2 Da), and collision energy 26 eV); and GSSG (m/z 307.0833, charge state +2, retention time 2.9 min, width 4 Da (+/– 2 Da), collision energy 20 eV).
Ion Mode:POSITIVE
Analysis Protocol File:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx