Return to study ST002795 main page

MB Sample ID: SA300153

Local Sample ID:Salvia-officinalis-subsp-lavandulifolia
Subject ID:SU002902
Subject Type:Plant
Subject Species:Salvia

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002902
Subject Type:Plant
Subject Species:Salvia

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
Salvia-officinalis-subsp-lavandulifoliaSA300153FL036430Salvia officinalis subsp lavandulifoliaTaxon

Collection:

Collection ID:CO002895
Collection Summary:Samples were collected from live mature plants and separated into root and leaf samples, with 3 biological replicates. Then samples were frozen in liquid nitrogen. The 2 parts of each plant were crushed into homogeneous powders respectively. In total, 71 species from three major distribution centers for this genus representing 8 out of the 11 recognized subclades of Salvia were sampled. To obtain sufficient materials for both chemical analyses, samples were collected at the vegetative or flowering stages.
Sample Type:Plant

Treatment:

Treatment ID:TR002911
Treatment Summary:No treatment, just species difference. In total, 71 species from three major distribution centers for this genus representing 8 out of the 11 recognized subclades of Salvia were sampled.

Sample Preparation:

Sampleprep ID:SP002908
Sampleprep Summary:All samples were ground into a fine powder, and 10 mg prepared powder was extracted with 1 mL of 70% methanol (v/v) containing warfarin (5 μg/mL) as the internal reference for 1 h in an ultrasonic bath (53 kHz, 350 W) at 4 °C. After centrifuging at 12,000 g at 4 °C for 30 min, the supernatant was used for UHPLC-QTOF-MS analysis.

Combined analysis:

Analysis ID AN004547
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity
Column Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI
MS instrument type QTOF
MS instrument name Waters Xevo-G2-XS
Ion Mode UNSPECIFIED
Units Normalized peak height

Chromatography:

Chromatography ID:CH003416
Instrument Name:Waters Acquity
Column Name:Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um)
Column Temperature:40 °C
Flow Gradient:The elution gradients were 98–80% A over 0–7 min, 80–78% A over 7–11 min, 78–40% A over 11–20 min, 40–35% A over 20–25 min, 25–28 min at 35% A, 35–5% A over 28–30 min, 30–33 min at 5% A, and final re-equilibration at 98% A for 5 min.
Flow Rate:0.40 mL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase

MS:

MS ID:MS004294
Analysis ID:AN004547
Instrument Name:Waters Xevo-G2-XS
Instrument Type:QTOF
MS Type:ESI
MS Comments:MS-DIAL v4.60 software. The data in the results table represent the average values for roots and leaves. Positive and negative ion-mode data were combined in the results table
Ion Mode:UNSPECIFIED
  logo