Return to study ST002856 main page

MB Sample ID: SA308799

Local Sample ID:Naive-NT3
Subject ID:SU002968
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU002968
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
Naive-NT3SA308799FL037178Vehicle (DMSO)Treatment

Collection:

Collection ID:CO002961
Collection Summary:Vehicle- and Enzalutamide-treated 16D cells were cultured with 30nM IACS-010759 for 24 hours
Sample Type:Cultured cells

Treatment:

Treatment ID:TR002977
Treatment Summary:Vehicle- and Enzalutamide-treated 16D cells were cultured with 30nM IACS-010759 in U13C-glucose-containing media for 24 hours

Sample Preparation:

Sampleprep ID:SP002974
Sampleprep Summary:Cell plates were placed on ice and quickly washed with ice-cold 0.9% (w/v) NaCl. Cells were immediately treated with 500 μL of ice-cold MeOH and 200 μL water containing 1 μg of the internal standard norvaline. Cells were then scraped and placed in 1.5 mL centrifuge tubes kept on ice. Next, 500 μL of chloroform was added, after which samples were vortexed for 1 minute and then spun at 10,000g for 5 minutes at 4C. The aqueous layer was transferred to a GC/MS sample vial and dried overnight using a refrigerated CentriVap.

Combined analysis:

Analysis ID AN004678
Analysis type MS
Chromatography type GC
Chromatography system Agilent 7890A
Column Agilent DB5-MS (30m x 0.25mm, 0.25um)
MS Type EI
MS instrument type GC-TOF
MS instrument name Agilent 5977B
Ion Mode POSITIVE
Units abundance

Chromatography:

Chromatography ID:CH003520
Instrument Name:Agilent 7890A
Column Name:Agilent DB5-MS (30m x 0.25mm, 0.25um)
Column Temperature:23
Flow Gradient:N/A (GC-MS)
Flow Rate:1 ml/min
Solvent A:N/A (GC-MS)
Solvent B:N/A (GC-MS)
Chromatography Type:GC

MS:

MS ID:MS004425
Analysis ID:AN004678
Instrument Name:Agilent 5977B
Instrument Type:GC-TOF
MS Type:EI
MS Comments:Data was analyzed using Agilent MassHunter software. Stable isotope tracing data was corrected for natural abundance of heavy isotopes with FluxFix software using a reference set of unlabeled metabolite standards2.
Ion Mode:POSITIVE
  logo