Return to study ST002973 main page

MB Sample ID: SA323088

Local Sample ID:SH_10
Subject ID:SU003086
Subject Type:Other abiotic sample

Select appropriate tab below to view additional metadata details:


Subject:

Subject ID:SU003086
Subject Type:Other abiotic sample

Factors:

Local Sample IDMB Sample IDFactor Level IDLevel ValueFactor Name
SH_10SA323088FL038544pre-filterTreatment
SH_10SA323088FL0385441Batch

Collection:

Collection ID:CO003079
Collection Summary:Two independently made batches of bacteria media was used for this study. For each batch, five aliquots from the same batch were used as replicates. Each aliquot was then split into halves for metabolite extraction. Following extraction, the one half was used as the pre-filtered controls and the other half was used for post-filtered sample that passed through the 0.2 micron filter membrane.
Sample Type:bacterial media

Treatment:

Treatment ID:TR003095
Treatment Summary:Metabolites extracted from mega medium, a rich and undefined bacterial medium, are filtered using a 96-well 0.2 micron filter plate. Here we compare the detection of metabolites in pre-filtered vs. post-filtered conditions from the same replicate, and five replicates are used for each of the two independent batches of media tested.

Sample Preparation:

Sampleprep ID:SP003092
Sampleprep Summary:The sample preparation procedure is described in detail in our preprint of this metabolomic protocol: https://protocolexchange.researchsquare.com/article/pex-2055/v1

Combined analysis:

Analysis ID AN004882
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Orbitrap Exploris 240
Ion Mode POSITIVE
Units raw ion count

Chromatography:

Chromatography ID:CH003684
Chromatography Summary:A published C18 reveres phase method was implemented with minor modifications. The C18 positive method (ESI+) used mobile phase solvents (LC-MS grade) consisting of 0.1% formic acid (Fisher) in water (A) and 0.1% formic acid in methanol (B). The gradient profile was from 0.5% B to 70% B in 4 minutes, from 70% B to 98% B in 0.5 minutes, and holding at 98% B for 0.9 minute before returning to 0.5% B in 0.2 minutes. The flow rate was 350 µL per minute. The sample injection volume was 5 µL. LC separations were made at 40C on separate columns fitted with a Vanguard pre-column of the same composition: Waters Acquity BEH 1.7 µm particle size, 2.1 mm id x 100 mm length (C18). Data were collected at a mass range of 70-1000 m/z at an acquisition rate of 2 spectra per second. Specific ion source parameters included Fragmentor (140V), Gas Temp (250oC), Sheath Gas Temp (200oC), and VCap (4000V).
Instrument Name:Thermo Vanquish
Column Name:Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um)
Column Temperature:40C
Flow Gradient:From 0.5% B to 70% B in 4 minutes, from 70% B to 98% B in 0.5 minutes, and holding at 98% B for 0.9 minute before returning to 0.5% B in 0.2 minutes.
Flow Rate:0.350 mL/minute
Solvent A:100% water + 0.1% formic acid
Solvent B:100% methanol + 0.1% formic acid
Chromatography Type:Reversed phase

MS:

MS ID:MS004626
Analysis ID:AN004882
Instrument Name:Thermo Orbitrap Exploris 240
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:Please see step-by-step details in our preprint for this metabolomic protocol: https://protocolexchange.researchsquare.com/article/pex-2055/v1
Ion Mode:POSITIVE