Return to study ST000543 main page

MB Sample ID: SA028103

Local Sample ID:R3C-180
Subject ID:SU000565
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Species Group:Human

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000825
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Dionex Ultimate 3000
Column Zorbax Extended C18
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap
Ion Mode POSITIVE
Units mole percent enrichment

Chromatography:

Chromatography ID:CH000589
Chromatography Summary:UPLC separations were performed with a Dionex UltiMate 3000 system (ThermoFisher Scientific) and a Zorbax Extended C18 column (Agilent; 5 cm × 2.1 mm, 1.8 µm). Solvent A was 99% water (Fisher Scientific), 1% acetonitrile (Fisher Scientific) with 0.1% formic acid (SigmaAldrich), and solvent B was 99% acetonitrile, 1% water and 0.1% formic acid. The gradient was as follows: 0-6 min: 5-20% B; 6-10 min: 20-95% B; 10-12 min: 95% B; 12-13 min: 95-5% B, at a flow rate of 0.3 mL min−1. An injection volume of 5 µL was used. The UPLC was connected to the ion source through the diverter valve. The HESI ion source was operated with +3.5 kV spray voltage and probe heater temperature of 300 °C. Sheath, auxiliary and spare gas were 47.5, 11.25 and 1.0 (arb. units), respectively. The capillary temperature was maintained at 275 °C.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Zorbax Extended C18
Flow Gradient:0-6 min: 5-20% B; 6-10 min: 20-95% B; 10-12 min: 95% B; 12-13 min: 95-5% B
Flow Rate:0.3ml/min
Solvent A:99% water/1% acetonitrile; 0.1% formic acid
Solvent B:99% acetonitrile/1% water; 0.1% formic acid
Chromatography Type:Reversed phase