Return to study ST000898 main page

MB Sample ID: SA052779

Local Sample ID:Control_1.3
Subject ID:SU000935
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Cell Biosource Or Supplier:ATCC
Cell Strain Details:Daoy (ATCC® HTB-186™)
Subject Comments:cancer cell line
Species Group:Human

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001460 AN001461
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Waters Acquity I-Class Waters Acquity I-Class
Column Waters Acquity BEH Amide (50 x 2.1mm,1.7um) Waters Acquity BEH Amide (50 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Waters Synapt G2 QTOF Waters Synapt G2 QTOF
Ion Mode POSITIVE NEGATIVE
Units Peak Area Peak Area

Chromatography:

Chromatography ID:CH001027
Chromatography Summary:The samples were analyzed on a Acquity UPLC I-class system from Waters. A HILIC-Amide column (1.7 μm, i.d. 2.1x50 mm) from Waters was used for sample separation and a non-linear gradient elution profile from 100 % A to 100 % B was used. Mobile phase A consisted of 90:10 acetonitrile/water with 10 mM ammonium formate and 0.1% FA while mobile phase B consisted of 50:50 acetonitrile/ water with 10 mM ammonium formate and 0.1% FA. The flow rate was 0.3 ml/min, the column temperature 40°C and the injection volume 5 µL.
Instrument Name:Waters Acquity I-Class
Column Name:Waters Acquity BEH Amide (50 x 2.1mm,1.7um)
Column Temperature:40°C
Flow Gradient:Non-linear gradient elution profile from 100 % A to 100 % B was used. In detail; 100% A was kept for 0.5 min then decreased non-linearly (slope-factor 8 in MassLynx) over 12.5 min to 100% B, 100% B was held for 3 min followed by 7 min at 100 % A to re-equilibrate the column for a total run-time of 23 min.
Flow Rate:0.3 ml/min
Injection Temperature:4°C
Sample Injection:5 μl
Solvent A:90% acetonitrile/10% water; 0.1% formic acid; 10 mM ammonium formate
Solvent B:50% acetonitrile/50% water; 0.1% formic acid; 10 mM ammonium formate
Analytical Time:23 min total sample run-time
Randomization Order:All samples were analyzed in a randomized order with three QC injections interspaced every eleventh injection
Chromatography Type:HILIC