Return to study ST000973 main page

MB Sample ID: SA058656

Local Sample ID:S-393
Subject ID:SU001012
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Human Ethnicity:Yoruba
Human Inclusion Criteria:apparent health, no recent treatment for any disease, no recent use of antibiotics, antimalarial or other drugs

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001594
Analysis type MS
Chromatography type Reversed phase
Chromatography system Q-Exactive hybrid quadrupole orbitrap mass spectrometer
Column Hypersil Gold
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units abundance of specific m/z (amu)

Chromatography:

Chromatography ID:CH001119
Chromatography Summary:Both urine and plasma samples were prepared using chilled methanol:water mixture (4:1). Separation was achieved with 150mm x 2.1mm, 1.9μ Hypersil Gold column with a 5μl injection volume. Gradient elution was performed using 0.1% formic acid in water (A) and acetonitrile (B) as mobile phase after optimization. The column was washed between each sample for stability and to eliminate carry-over. The flow rate was 0.35 ml/min. MS acquisition was performed on the Q-Exactive hybrid quadrupole orbitrap mass spectrometer (ThermoFisher, MA, USA) operated in positive and negative electrospray ionization (ESI) modes. The scan range was 70 to 1000 m/z.
Instrument Name:Q-Exactive hybrid quadrupole orbitrap mass spectrometer
Column Name:Hypersil Gold
Flow Rate:0.35ml/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase