Return to study ST001216 main page
MB Sample ID: SA085918
Local Sample ID: | BALF_17 |
Subject ID: | SU001283 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Gender: | Male and female |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002027 | AN002028 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | HILIC |
Chromatography system | Agilent 6545 | Agilent 6520 |
Column | Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) | Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um) |
MS Type | ESI | ESI |
MS instrument type | QTOF | QTOF |
MS instrument name | Agilent 6545 QTOF | Agilent 6520 QTOF |
Ion Mode | POSITIVE | POSITIVE |
Units | Abundance (Log2) | Abundance (Log2) |
Chromatography:
Chromatography ID: | CH001468 |
Chromatography Summary: | Reversed phase samples from the lipid fraction were randomized in the worklist and run randomly in triplicate using an Agilent 1290 series pump with an Agilent Zorbax Rapid Resolution HD (RRHD) SB-C18, 1.8 micron (2.1 × 100 mm) analytical column and an Agilent Zorbax SB-C18, 1.8 micron (2.1 × 5 mm) guard column. The autosampler tray temperature was set at 4 °C, column temperature was set at 60 °C, and the sample injection volume was 8 µL for BAL and 4 µL for plasma. The flow rate was 0.7 mL/min with the following mobile phases: mobile phase A was water with 0.1% formic acid, and mobile phase B was 60:36:4 isopropyl alcohol:acetonitrile:water with 0.1% formic acid. Gradient elution was as follows: 0–0.5 minutes 30–70% B, 0.5–7.42 minutes 70–100% B, 7.42–10.4 minutes 100% B, 10.4–10.5 minutes 100–30% B, 10.5–15.1 minutes 30% B. |
Instrument Name: | Agilent 6545 |
Column Name: | Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) |
Column Temperature: | 60 |
Flow Gradient: | 0-0.5 minutes 30-70% B, 0.5-7.42 minutes 70-100% B, 7.42-10.4 minutes 100% B, 10.4-10.5 minutes 100-30% B, 10.5-15.1 minutes 30% B |
Flow Rate: | 0.7 mL/min |
Solvent A: | 100% water; 0.1% formic acid, |
Solvent B: | 60% isopropanol/36% acetonitrile/4% water; 0.1% formic acid |
Chromatography Type: | Reversed phase |
Chromatography ID: | CH001469 |
Chromatography Summary: | The samples from the aqueous small molecule fraction were analyzed randomly in triplicate using an Agilent 1290 series pump using a Phenomenex Kinetex HILIC, 2.6 µm, 100 Å (2.1 × 50 mm) analytical column and an Agilent Zorbax Eclipse Plus-C8 5 µm (2.1 × 12.5 mm) narrow bore guard column. The autosampler tray temperature was set at 4 °C, column temperature was set at 20 °C, and the sample injection volume was 1 µL for both BAL and plasma. The flow rate of 0.6 mL/min with the following mobile phases: mobile phase A was 50% ACN with pH 5.8 ammonium acetate, and mobile phase B was 90% ACN with pH 5.8 ammonium acetate. Gradient elution was as follows: 0.2 minutes 100% B, 0.2–2.1 minutes 100–90% B, 2.1–8.6 minutes 90–50% B, 8.6–8.7 minutes 50–0% B, 8.7–14.7 minutes 0% B, 14.7–14.8 minutes 0–100% B, 14.8–24.8 minutes 100% B. |
Instrument Name: | Agilent 6520 |
Column Name: | Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um) |
Column Temperature: | 20 |
Flow Gradient: | 0.2 minutes 100% B, 0.2-2.1 minutes 100-90% B, 2.1-8.6 minutes 90-50% B, 8.6-8.7 minutes 50-0% B, 8.7-14.7 minutes 0% B, 14.7-14.8 minutes 0-100% B, 14.8-24.8 minutes 100% B. |
Flow Rate: | 0.6 mL/min |
Solvent A: | 50% acetonitrile/50% water; ammonium acetate, pH 5.8 |
Solvent B: | 90% acetonitrile/5% water, pH 5.8 |
Chromatography Type: | HILIC |