MB Sample ID: SA086276

Local Sample ID:BBSe-1
Subject ID:SU001288
Subject Type:Invertebrate
Subject Species:Bombus impatiens
Taxonomy ID:132113

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002034
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity I-Class
Column Waters Acquity CSH Phenyl-Hexyl (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type QTOF
MS instrument name Waters Synapt G2 Si QTOF
Ion Mode POSITIVE
Units Peak Area

Chromatography:

Chromatography ID:CH001474
Chromatography Summary:We used a Synapt G2-Si quadrupole time-of-flight mass spectrometer (Waters, Milford, MA) coupled to an I-class UPLC system (Waters) for LC-MS analyses in the UC Riverside Metabolomics Core Facility. We carried out separations on a CSH phenyl-hexyl column (2.1 x 100 mm, 1.7 µM) (Waters, Milford, MA), with the following mobile phases: A. Water with 0.1% formic acid and B. Acetonitrile with 0.1% formic acid at a flow rate of 250 µL/min at 40° C. We injected 2 µL of sample extract, and the gradient was as follows: 0 min, 1% B; 1 min, 1% B; 8 min, 40% B; 24 min, 100% B; 26.5 min, 100% B; 27 min, 1% B.
Methods Filename:jroth002_LC_method.docx
Instrument Name:Waters Acquity I-Class
Column Name:Waters Acquity CSH Phenyl-Hexyl (100 x 2.1mm,1.7um)
Column Temperature:40
Flow Gradient:0 min, 1% B; 1 min, 1% B; 8 min, 40% B; 24 min, 100% B; 26.5 min, 100% B; 27 min, 1% B.
Flow Rate:250 µL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase