Return to study ST001406 main page

MB Sample ID: SA114153

Local Sample ID:POTR_03_sAT
Subject ID:SU001480
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:11-20 years
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002349
Analysis type MS
Chromatography type GC
Chromatography system Thermo Trace 1310
Column Agilent DB5-MS (30m x 0.25mm, 0.25um)
MS Type EI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive GC Orbitrap GC-MS/MS
Ion Mode POSITIVE
Units pg/g

Chromatography:

Chromatography ID:CH001721
Chromatography Summary:Tissue extracts were analyzed using a Thermo Scientific 1310 gas chromatograph connected to a Q Exactive GC Orbitrap GC-MS/MS ultra-high-resolution mass spectrometer and Triplus RSH autosampler. A 2 µL aliquot of extract was injected into an inlet maintained at 250ºC in pulsed split-less mode. The analytes were separated on an Agilent DB-5MSUI capillary column (30m length × 0.25mm inner diameter × 0.25µm film thickness) using high purity helium (99.999% purity) as the carrier gas at a constant flow rate of 1 mL/min. The oven temperature program consisted of an initial temperature of 100ºC for 1 min, increased to 180ºC at 25ºC/min; followed by a temperature ramp to 215ºC at 5ºC/min, and finally increased to 300ºC at 25ºC/min and held for 10 min, resulting in a total run time of 26.6 min.
Instrument Name:Thermo Trace 1310
Column Name:Agilent DB5-MS (30m x 0.25mm, 0.25um)
Flow Rate:1 mL/min
Injection Temperature:250C
Internal Standard:[13C6]-Anthracene, [13C12]-PCB28, [DIETHYL-D10]-Chlorpyrifos, [13C12]-PCB101, [13C12]-4,4'-DDE, [13C12]-PCB153, [13C12]-PCB180, [13C12]-PBDE47, [13C10]-Mirex, [13C6]-cis-Permethrin, [13C12]-PBDE99 and [13C12]-PBB153
Sample Injection:2 uL
Analytical Time:26.6
Oven Temperature:The oven temperature program consisted of an initial temperature of 100ºC for 1 min, increased to 180ºC at 25ºC/min; followed by a temperature ramp to 215ºC at 5ºC/min, and finally increased to 300ºC at 25ºC/min and held for 10 min
Transferline Temperature:280
Sample Syringe Size:10uL