MB Sample ID: SA114204

Local Sample ID:q3June2014_1c
Subject ID:SU001481
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:11-20 years
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002350 AN002351
Analysis type MS MS
Chromatography type HILIC Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 Thermo Dionex Ultimate 3000
Column Waters XBridge BEH Amide XP HILIC (50 x 2.1mm,2.5um) Higgins endcapped C18 stainless steel (50 x 2.1mm,3um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Peak intensity Peak intensity

Chromatography:

Chromatography ID:CH001722
Chromatography Summary:The HILIC column is operated parallel to reverse phase column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10- port and 6-port switching valves. During operation of HILIC separation method, the MS is operated in positive ion mode and 10 μL of sample is injected onto the HILIC column while the reverse phase column is flushing with wash solution. Flow rate is maintained at 0.35 mL/min until 1.5 min, increased to 0.4 mL/min at 4 min and held for 1 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 2% formic acid (v/v) in LC-MS grade water. Initial mobile phase conditions are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time of 5 min. During the flushing phase (reverse phase analytical separation), the HILIC column is equilibrated with a wash solution of 77.5% A, 20% B, 2.5% C.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Waters XBridge BEH Amide XP HILIC (50 x 2.1mm,2.5um)
Column Temperature:60
Flow Gradient:Initial mobile phase conditions are 22.5% A, 75% B, 2.5% C hold for 1.5 min, with linear gradient to 77.5% A, 20% B, 2.5% C at 4 min, hold for 1 min, resulting in a total analytical run time of 5 min.
Flow Rate:0.35- 0.4 mL/min
Internal Standard:[13C6]-D-glucose, [15N,13C5]- L-methionine, [13C5]-L-glutamic acid, [15N]-L-tyrosine, [3,3-13C2]-cystine, [trimethyl- 13C3]-caffeine, [U-13C5, U-15N2]-L-glutamine
Sample Injection:10 uL
Solvent A:100% water(A), 100% acetonitrile(B), 100% water; 2% formic acid(C)
Solvent B:100% water(A), 100% acetonitrile(B), 100% water; 2% formic acid(C)
Analytical Time:5 min
Chromatography Type:HILIC
  
Chromatography ID:CH001723
Chromatography Summary:The C18 column is operated parallel to the HILIC column for simultaneous analytical separation and column flushing through the use of a dual head HPLC pump equipped with 10-port and 6- port switching valves. During operation of the C18 method, the MS is operated in negative ion mode and 10 μL of sample is injected onto the C18 column while the HILIC column is flushing with wash solution. Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min. Solvent A is 100% LC-MS grade water, solvent B is 100% LC-MS grade acetonitrile and solvent C is 10mM ammonium acetate in LC-MS grade water. Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min. During the flushing phase (HILIC analytical separation), the C18 column is equilibrated with a wash solution of 0% A, 95% B, 5% C until 2.5 min, followed by an equilibration solution of 60% A, 35% B, 5% C for 2.5 min.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Higgins endcapped C18 stainless steel (50 x 2.1mm,3um)
Column Temperature:60
Flow Gradient:Initial mobile phase conditions are 60% A, 35% B, 5% C hold for 0.5 min, with linear gradient to 0% A, 95% B, 5% C at 1.5 min, hold for 3.5 min, resulting in a total analytical run time of 5 min.
Flow Rate:Flow rate is maintained at 0.4 mL/min until 1.5 min, increased to 0.5 mL/min at 2 min and held for 3 min.
Internal Standard:[13C6]-D-glucose, [15N,13C5]- L-methionine, [13C5]-L-glutamic acid, [15N]-L-tyrosine, [3,3-13C2]-cystine, [trimethyl- 13C3]-caffeine, [U-13C5, U-15N2]-L-glutamine
Sample Injection:10 uL
Solvent A:100% water(A), 100% acetonitrile(B), 100%water; 10mM ammonium acetate(C)
Solvent B:100% water(A), 100% acetonitrile(B), 100%water; 10mM ammonium acetate(C)
Analytical Time:5 min
Chromatography Type:Reversed phase