Return to study ST001662 main page

MB Sample ID: SA152156

Local Sample ID:mv08618_47.mzXML
Subject ID:SU001739
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Pooled

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002713
Analysis type MS
Chromatography type Reversed phase
Chromatography system Agilent 6550
Column Waters Acquity BEH C18 (100 x 2mm,1.7um)
MS Type EI
MS instrument type QTOF
MS instrument name Agilent 6550 QTOF
Ion Mode POSITIVE
Units peak area

Chromatography:

Chromatography ID:CH002002
Chromatography Summary:The metabolomic analysis was carried out using an Ultra-Performance Liquid Chromatography (UPLC) system coupled to an iFunnel Q-ToF (quadrupole-time-of-flight) Agilent 6550 mass spectrometer (Agilent Technologies, CA, USA) using an UPLC BEH C18 (100 x 2.1 mm, 1.7 µm, Waters, Wexford, Ireland) column from Waters (Wexford, Ireland). Autosampler and column temperatures were set to 4 °C and 40 °C, respectively, and the injection volume was 5 µl. Mobile phase A and mobile phase B consisted of H2O and acetonitrile, both containing 0.1% of formic acid. A gradient elution was performed at a flow rate of 400 µl min-1 along 14 min as follows: initial conditions of 98% of mobile phase A was maintained for 1 min, and then decreased until 75% in 2 min, 50% in 3 min and 5% in 3 more min. 95% of mobile phase B was held for 3 min and then, a 0.55 min gradient was used to return to the initial conditions, which were held for 2.5 min to totally column recovery.
Instrument Name:Agilent 6550
Column Name:Waters Acquity BEH C18 (100 x 2mm,1.7um)
Column Temperature:40
Flow Gradient:initial conditions of 98% of mobile phase A was maintained for 1 min, and then decreased until 75% in 2 min, 50% in 3 min and 5% in 3 more min. 95% of mobile phase B was held for 3 min and then, a 0.55 min gradient was used to return to the initial conditions, which were held for 2.5 min to totally column recovery
Flow Rate:400ul/min
Solvent A:100% water; 0.1% formic acid
Solvent B:acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase