Return to study ST001681 main page

MB Sample ID: SA154771

Local Sample ID:091_38_C
Subject ID:SU001758
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002742 AN002743 AN002744 AN002745
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) Agilent Zorbax SBaq (50 x 2.1mm,1.7um) Agilent Zorbax SBaq (50 x 2.1mm,1.7um)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive HF hybrid Orbitrap Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units MS Counts MS Counts MS Counts MS Counts

Chromatography:

Chromatography ID:CH002026
Chromatography Summary:HILIC experiments were performed using a ZIC-HILIC column 2.1x100 mm, 3.5μm, 200Å (Merck Millipore) and mobile phase solvents consisting of 10mM ammonium acetate in 50/50 acetonitrile/water (A) and 10 mM ammonium acetate in 95/5 acetonitrile/water (B).(Contrepois et al., 2015)
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
Column Temperature:40
Flow Rate:0.5 ml/min
Solvent A:95% acetonitrile/5% water; 10 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 10 mM ammonium acetate
Chromatography Type:HILIC
  
Chromatography ID:CH002027
Chromatography Summary:RPLC experiments were performed using a Zorbax SBaq column 2.1 x 50 mm, 1.7 μm, 100Å (Agilent Technologies) and mobile phase solvents consisting of 0.06% acetic acid in water (A) and 0.06% acetic acid in methanol (B). (Contrepois et al., 2015)
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Agilent Zorbax SBaq (50 x 2.1mm,1.7um)
Column Temperature:60
Flow Rate:0.6 ml/min
Solvent A:100% water; 0.06% acetic acid
Solvent B:100% methanol; 0.06% acetic acid
Chromatography Type:Reversed phase