Return to study ST001734 main page

MB Sample ID: SA162669

Local Sample ID:POL-04
Subject ID:SU001811
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002823
Analysis type MS
Chromatography type HILIC
Chromatography system Agilent 1260
Column Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name Agilent 6470 QQQ
Ion Mode NEGATIVE
Units ug/mg

Chromatography:

Chromatography ID:CH002087
Chromatography Summary:We used HPLC system (1260 Infinity, Agilent Technologies). Metabolites were separated on a Kinetex hydrophilic interaction liquid chromatography (HILIC) silica column (150 mm × 2.1 mm, particle size 100Å, Phenomenex, USA) maintained at 25 ºC. The mobile phases consisted of A) water, and B) acetonitrile, both with 7.5 mM ammonium acetate and 0.1% acetic acid, with a final pH of 4.0 in the aqueous phase. The flow rate was 0.5 mL/min with an injection volume of 5 µl. Gradient started with 5% of A for 2 min, then increased up to 50% until 12 min, and back to initial conditions until 22 min.
Instrument Name:Agilent 1260
Column Name:Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
Column Temperature:25ºC
Flow Rate:0.5 mL/min
Solvent A:100% water; 0.1% acetic acid; 7.5 mM ammonium acetate, pH 4
Solvent B:100% acetonitrile; 0.1% acetic acid; 7.5 mM ammonium acetate
Chromatography Type:HILIC