MB Sample ID: SA164964

Local Sample ID:S12R2
Subject ID:SU001853
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002883
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000 RS
Column Thermo Accucore HILIC (100 x 2.1mm,2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units arbitray unit

Chromatography:

Chromatography ID:CH002138
Chromatography Summary:The UHPLC system was equipped with a binary pump, an autosampler and a column thermostat. The autosampler was set at 8 °C. The column oven temperature was 30 °C. Mobile phase A was 98% acetonitrile, 2% water and 0.1% formic acid while mobile phase B was 98% water, 2% acetonitrile, 30 mM ammonium formate and 0.1% formic acid. The mobile phase was freshly prepared before use. The optimized stepwise linear gradient (10% B in the first 2 min, 10% - 30% B in the next 5 min, followed by 30% - 60% in 3 min, and finally 60% - 90% in 5 min.
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Thermo Accucore HILIC (100 x 2.1mm,2.6um)
Column Temperature:25
Solvent A:98% acetonitrile/2% water; 0.1% formic acid
Solvent B:98% water/2% acetonitrile; 0.1% formic acid; 30 mM ammonium formate
Sample Loop Size:10
Chromatography Type:HILIC