Return to study ST002046 main page

MB Sample ID: SA192411

Local Sample ID:exp6_hilic_neg_02
Subject ID:SU002128
Subject Type:Invertebrate
Subject Species:Caenorhabditis elegans
Taxonomy ID:6239

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003330 AN003331 AN003332 AN003333
Analysis type MS MS MS MS
Chromatography type Reversed phase Reversed phase HILIC HILIC
Chromatography system Thermo Vanquish Thermo Vanquish Thermo Vanquish Thermo Vanquish
Column Thermo Scientific Accucore C30 (150 x 2.1mm,2.6um) Thermo Scientific Accucore C30 (150 x 2.1mm,2.6um) Waters Acquity BEH Amide (150 x 2.1mm,1.7um) Waters Acquity BEH Amide (150 x 2.1mm,1.7um)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap Thermo Q Exactive HF-X Orbitrap Thermo Q Exactive HF-X Orbitrap Thermo Q Exactive HF-X Orbitrap
Ion Mode POSITIVE NEGATIVE POSITIVE NEGATIVE
Units Peak height Peak height Peak height Peak height

Chromatography:

Chromatography ID:CH002467
Chromatography Summary:The following gradient program was used: -5.0 20% B; 0.0 min 20% B; 1.0 min 60% B; 5.0 min 70% B; 5.5 min 85% B; 8.0 min 90% B; 8.2-10.5 min 100% B; 10.7-12.0 min 20% B. A curve 5 value was set for -5.0 and 0.0 minutes, and a curve 6 for the remainder of the gradient.
Instrument Name:Thermo Vanquish
Column Name:Thermo Scientific Accucore C30 (150 x 2.1mm,2.6um)
Column Temperature:55
Flow Rate:0.400 mL/min
Sample Injection:2.0 uL
Solvent A:40% water/60% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase
  
Chromatography ID:CH002468
Chromatography Summary:The following gradient program was used: -5.0 min 95% B; 0.0-0.5 min 95% B; 8.0-9.4 min 40% B; 9.5-11.0 min 95% B. A curve 5 value was set for -5.0 and 0.0 minutes, a curve 6 at 0.5 min, curve 7 at 8.0 min, and a curve 6 for the remainder of the gradient.
Instrument Name:Thermo Vanquish
Column Name:Waters Acquity BEH Amide (150 x 2.1mm,1.7um)
Column Temperature:40
Flow Rate:0.400 mL/min
Sample Injection:2.0 uL
Solvent A:80% water/20% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:100% acetonitrile; 0.1% formic acid