Return to study ST002201 main page
MB Sample ID: SA211314
Local Sample ID: | Tsu-0_W-N-_P3ID34_a |
Subject ID: | SU002287 |
Subject Type: | Plant |
Subject Species: | Arabidopsis thaliana |
Taxonomy ID: | 3702 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003604 |
---|---|
Analysis type | MS |
Chromatography type | GC |
Chromatography system | Agilent 7890B |
Column | Restek Rxi-5Sil (30m x 0.25mm,0.25m) with 10m precolumn |
MS Type | EI |
MS instrument type | Single quadrupole |
MS instrument name | Agilent 5977B |
Ion Mode | POSITIVE |
Units | µg/mg FW and arbitrary/mg FW |
Chromatography:
Chromatography ID: | CH002664 |
Chromatography Summary: | The instrument was an Agilent 7890B gas chromatograph coupled to an Agilent 5977B mass spectrometer. The column was an Rxi-5SilMS from Restek (30 m with 10 m integraguard column). The liner (Restek # 20994) was changed before the analysis. Oven temperature ramp was 70 °C for 7 min then 10 °C/min to 330 °C for 5 min (run length 38 min). Helium constant flow was 0.7 mL/min. Temperatures were the following: injector: 250°C, transfer line: 290°C, source: 250 °C and quadripole 150 °C. 5 scans per second were acquired spanning a 50 to 600 Da range. Instrument was tuned with PFTBA with the 69 m/z and 219 m/z of equal intensities. 5 scans per second were acquired. The split mode conditions were: 70°C for 2 min then 30°C per min to 330 °C for 5 min. Helium constant flow 1 mL/min. |
Instrument Name: | Agilent 7890B |
Column Name: | Restek Rxi-5Sil (30m x 0.25mm,0.25m) with 10m precolumn |
Chromatography Type: | GC |