Return to study ST002240 main page

MB Sample ID: SA214093

Local Sample ID:0_GAT_2-pos
Subject ID:SU002326
Subject Type:Plant
Subject Species:Arabidopsis thaliana
Taxonomy ID:3702
Age Or Age Range:10 day old seedlings
Species Group:Roots

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003657 AN003658
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Agilent 1290 Infinity II Agilent 1290 Infinity II
Column SeQuant ZIC-HILIC (100 x 2.1mm,3.5um) SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Intensity Intensity

Chromatography:

Chromatography ID:CH002710
Chromatography Summary:Chromatographic separation using HILIC was achieved using an Agilent 1290 Infinity II UHPLC system equipped with a SeQuant® ZIC®-HILIC (100 x 2.1 x 3.5 µm) column (EMD Millipore, Burlington, MA). Mobile phases A and B were comprised of 5 mM ammonium acetate (pH 4.0) in water and 90% acetonitrile with 0.1 % acetic acid, respectively. A flow rate of 0.3 mL min-1 was used to elute compounds with the following gradient: 87% B for 5 minutes, decreased to 55% B over the next 8 minutes and held for 2.5 minutes before returning to 87% and equilibrating the column for 3 minutes.
Instrument Name:Agilent 1290 Infinity II
Column Name:SeQuant ZIC-HILIC (100 x 2.1mm,3.5um)
Column Temperature:40
Flow Rate:0.3 mL min-1
Internal Standard:Equisplash
Solvent A:100% water; 5 mM ammonium acetate, pH 4
Solvent B:90% acetonitrile/10% water; 0.1% acetic acid
Chromatography Type:HILIC
  logo