Return to study ST002273 main page

MB Sample ID: SA217814

Local Sample ID:A03-01_59_1_1701
Subject ID:SU002359
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Not applicable
Cell Biosource Or Supplier:Sigma-Aldrich (St. Louis, MO)
Cell Strain Details:Hep3B

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003715
Analysis type MS
Chromatography type Reversed phase
Chromatography system UHPLC-QTOF-MS
Column Hamilton Intensity Solo 2 C18 (100 mm x 2.1 mm,1.8um)
MS Type ESI
MS instrument type Trapped Ion Mobility Q-TOF
MS instrument name Bruker timsTOF
Ion Mode POSITIVE
Units AU

Chromatography:

Chromatography ID:CH002752
Chromatography Summary:The Elute UHPLC and Q-TOF Mass Spectrometer (Bruker, Bremen, Germany) were utilized for metabolite and peptide detection. The Elute HPG 1300 pumps, Elute Au-tosampler (Bruker, Bremen, Germany), and Hamilton® Intensity Solo 2 C18 column (100 mm x 2.1 mm, 1.8 m beads) were employed using reversed-phase chromatog-raphy. Solvents used for separation were 0.1 % FA in LC grade water (solvent A) and 0.1 % FA in ACN (solvent B). Each metabolite and protein extract were analyzed in duplicate. For metabolomics analysis, the column was kept at 35°C, and each sample was injected twice with an in-jection volume of 10 µL. Sample elution was performed in 30 min gradient starting with 1% ACN for 2 min and then ramped to 99% ACN within 15 min. After that, 99% ACN was kept for 3 min, and then the re-equilibration to 1% ACN was done for 10 min. The flow rate was 0.25 mL/min for 20 min and then 0.35 mL/min for 8.3 min and then the flow rate set at 0.25 mL/min for 1.7 min.
Instrument Name:UHPLC-QTOF-MS
Column Name:Hamilton Intensity Solo 2 C18 (100 mm x 2.1 mm,1.8um)
Column Temperature:35°C
Flow Gradient:Sample elution was performed in 30 min gradient starting with 1% ACN for 2 min and then ramped to 99% ACN within 15 min. After that, 99% ACN was kept for 3 min, and then the re-equilibration to 1% ACN was done for 10 min
Flow Rate:was 0.25 mL/min for 20 min and then 0.35 mL/min for 8.3 min and then the flow rate set at 0.25 mL/min for 1.7 min.
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase
  logo