MB Sample ID: SA244206

Local Sample ID:SC_20201109_HILICp_FMS_CSF_D02_C_1_1
Subject ID:SU002531
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003979
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column Waters ACQUITY UPLC BEH HILIC (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units time_m/z

Chromatography:

Chromatography ID:CH002942
Chromatography Summary:CSF extracts (5uL injection volume) were separated on an ACQUITY UPLC BEH Amide HILIC 1.7μm, 2.1 × 100 mm column (Waters Corporation, Milford, MA) held at 30°C as previously described (cite). Briefly, liquid chromatography was performed at 200 µL min−1 using solvent A (5 mM ammonium formate in 90% water, 10% acetonitrile and 0.1% formic acid) and solvent B (5 mM ammonium formate in 90% acetonitrile, 10% water and 0.1% formic acid) with a gradient length of 30 min.
Methods Filename:Metabolomics_Analyses.pdf
Instrument Name:Thermo Vanquish
Column Name:Waters ACQUITY UPLC BEH HILIC (100 x 2.1mm,1.7um)
Column Temperature:30
Flow Gradient:30 min
Flow Rate:0.20mL/min
Solvent A:90% water, 10% acetonitrile, 5mM Ammonium Formate, 0.1%FA
Solvent B:10% water, 90% acetonitrile, 5mM Ammonium Formate, 0.1%FA
Chromatography Type:HILIC