Return to study ST002778 main page

MB Sample ID: SA297431

Local Sample ID:2021-02-01 JL10 SO2 MRM LM TR2_1-C
Subject ID:SU002885
Subject Type:Amphibian
Subject Species:Xenopus laevis
Taxonomy ID:8355

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004522 AN004523 AN004524
Analysis type MS MS MS
Chromatography type Ion exchange Ion exchange HILIC
Chromatography system Waters ACQUITY I-Class Waters Acquity I-Class Waters Acquity I-Class
Column Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um) Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um) ACQUITY UPLC BEH amide (100 x 1mm, 1.7um)
MS Type ESI ESI ESI
MS instrument type QTOF QTOF QTOF
MS instrument name Bruker timsTOF PRO Bruker timsTOF PRO Bruker timsTOF PRO
Ion Mode NEGATIVE NEGATIVE POSITIVE
Units Counts Counts Counts

Chromatography:

Chromatography ID:CH003396
Chromatography Summary:A 2.5 µL of the metabolite extract was loaded onto a reversed-phase LC column featuring anion-exchange chemistry (Atlantis PREMIER BEH C18 AX Column, Waters, 1.7 µm, 2.1 × 100 mm) and separated at 500 uL/min in a micro-flow LC system (ACQUITY I-Class, Waters) using a gradient of buffer A (100% ACN) and buffer B (95% water, 5% ACN with 1 mM ammonium acetate; pH 7.8) as follows: 5% was held 0–¬0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min. To recover and recondition the analytical column before the next sample injection, solvent B was then decreased to 5% over 2 min and the column was rinsed for 7 min. The column temperature was set to 30 °C.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters ACQUITY I-Class
Column Name:Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um)
Column Temperature:30
Flow Gradient:5% B was held 0–0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min
Flow Rate:500 uL/min
Solvent A:100% ACN
Solvent B:95% water/5% acetonitrile; 1 mM ammonium acetate; pH 7.8
Chromatography Type:Ion exchange
  
Chromatography ID:CH003397
Chromatography Summary:A 2.5 µL of the metabolite extract was loaded onto a reversed-phase LC column featuring anion-exchange chemistry (Atlantis PREMIER BEH C18 AX Column, Waters, 1.7 µm, 2.1 × 100 mm) and separated at 500 uL/min in a micro-flow LC system (ACQUITY I-Class, Waters) using a gradient of buffer A (100% ACN) and buffer B (95% water, 5% ACN with 1 mM ammonium acetate; pH 7.8) as follows: 5% was held 0–¬0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min. To recover and recondition the analytical column before the next sample injection, solvent B was then decreased to 5% over 2 min and the column was rinsed for 7 min. The column temperature was set to 30 °C.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters Acquity I-Class
Column Name:Waters Atlantis Premier BEH C18 AX (100 x 2.1mm, 1.7um)
Column Temperature:30
Flow Gradient:5% B was held 0–0.5 min, then linearly increased to 90% over 9.5 min, and held at 90% for 3 min
Flow Rate:500 uL/min
Solvent A:100% ACN
Solvent B:95% water/5% acetonitrile; 1 mM ammonium acetate; pH 7.8
Chromatography Type:Ion exchange
  
Chromatography ID:CH003398
Chromatography Summary:A 2-µL volume of the metabolite extract was loaded onto an ACQUITY UPLC BEH amide column (130 A, 1.7 µm, 1 mm × 100 mm) and separated at 45 ℃ using a mobile phase delivered at 130 uL/min in a micro-LC system (Waters ACQUITY I-Class). The separation was performed in buffer A (water containing 0.1% v/v FA) with a gradient of buffer B (100% ACN with 0.1% v/v FA) as follows: 1% A for 0.05 min, then ramped to 80% A over 3.20 min, held at 80% A for 2 min, ramped to 1% A in 1.75 min. To recover and condition the analytical column for the next sample injection, buffer A was held at 1% for 7 min.
Methods Filename:Baxi_2023_Organizer_Work_MeabolomicsWorkBench_Description_2023-07-02.docx
Instrument Name:Waters Acquity I-Class
Column Name:ACQUITY UPLC BEH amide (100 x 1mm, 1.7um)
Column Temperature:45
Flow Gradient:1% A for 0.05 min, then ramped to 80% A over 3.20 min, held at 80% A for 2 min, ramped to 1% A in 1.75 min
Flow Rate:130 uL/min
Solvent A:water containing 0.1% v/v FA
Solvent B:100% acetonitrile; 0.1% v/v FA
Chromatography Type:HILIC