Return to study ST003017 main page

MB Sample ID: SA327615

Local Sample ID:Sp64
Subject ID:SU003131
Subject Type:Mammal
Subject Species:Florida manatee (Trichechus manatus latirostris)
Age Or Age Range:Adult

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004950 AN004951
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Waters ACQUITY UPLC BEH C18(100 x 2.1mm,1.7um) Waters ACQUITY UPLC BEH C18(100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Peak area Peak area

Chromatography:

Chromatography ID:CH003735
Chromatography Summary:The extracted lipid samples, pooled groups, quality controls, and blanks were analyzed by UHPLC-MS/MS on a Thermo Vanquish UHPLC system coupled to a Thermo Q-Exactive Orbitrap mass spectrometer. The column used was a Waters Acquity BEH, 100 mm x 2.1 mm x 1.7 µm. Mobile phase A and B contained 60:40 acetonitrile/water (v/v) and 90:10 isopropanol/acetonitrile (v/v) respectively, with both phases containing 5mM ammonium formate and 0.1% formic acid. The mobile phase gradient was as follows: a linear increase of solvent B from 40% to 55% over 7 min, then increased to 65% by min 8, held at 65% for 4 min, linear increase to 95% over 8 min, increased to 100% over 2 min, solvent B was held at 100% for 5 min before decreasing to 40% in a tenth of a min. Finally, solvent B was held at 100% for another 3 min. The sample injection volume was 10 µL at a flow rate of 0.25 mL/min, the column compartment was 45 °C, and the autosampler temperature was 10 °C.
Instrument Name:Thermo Vanquish
Column Name:Waters ACQUITY UPLC BEH C18(100 x 2.1mm,1.7um)
Column Temperature:45
Flow Gradient:a linear increase of solvent B from 40% to 55% over 7 min, then increased to 65% by min 8, held at 65% for 4 min, linear increase to 95% over 8 min, increased to 100% over 2 min, solvent B was held at 100% for 5 min before decreasing to 40% in a tenth of a min. Finally, solvent B was held at 100% for another 3 min.
Flow Rate:0.25 mL/min
Solvent A:60:40 acetonitrile/water (v/v), 5mM ammonium formate and 0.1% formic acid
Solvent B:90:10 isopropanol/acetonitrile (v/v), 5mM ammonium formate and 0.1% formic acid
Chromatography Type:Reversed phase