Return to study ST003050 main page

MB Sample ID: SA331299

Local Sample ID:Donor_6_Plasma_after_Aspirin_2
Subject ID:SU003165
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005001
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column Phenomenex Kinetex XB-C18 (150 x 2.1mm, 2.6um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode NEGATIVE
Units normalized AUC

Chromatography:

Chromatography ID:CH003779
Chromatography Summary:For LC-MS analyses, analytes were separated using a Thermo Scientific Vanquish (UHPLC) system equipped with a Kinetex C18-column (2.6 µm, XB-C18, 100 A° , LC Column 150 * 2.1 mm; Phenomenex) applying a gradient flow profile (mobile phase A: H2O + 0.2% FA, mobile phase B: ACN:MeOH (vol% 90:10) + 0.2% FA) starting at 35% B and increasing to 90% B (1–10 min), further increasing to 99% B within 0.5 min and held for 5 min. Solvent B was then decreased to the initial level of 35% within 0.5 min and the column was equilibrated for 4 min, resulting in a total run time of 20 min. The flow rate was kept at 200 µL min-1 and the column oven temperature at 40°C. The injection volume was 20 µL and all samples were analysed in technical duplicates.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex XB-C18 (150 x 2.1mm, 2.6um)
Column Temperature:40
Flow Gradient:0min with 35% B to 90% B (1–10 min), further increasing to 99% B within 0.5 min and held for 5 min. Solvent B was then decreased to the initial level of 35% within 0.5 min and the column was equilibrated for 4 min, resulting in a total run time of 20 min.
Flow Rate:200 µL/min
Solvent A:100% water; 0.2% formic acid
Solvent B:90% acetonitrile/10% methanol; 0.2% formic acid
Chromatography Type:Reversed phase