Return to study ST003168 main page

MB Sample ID: SA343049

Local Sample ID:old_10
Subject ID:SU003287
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:Young donors (aged 25 years or younger) and older donors (aged 50 years or older).
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005198 AN005199
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Waters HPLC C18 (100 x 2.1mm, 1.7um) Waters HPLC C18 (100 x 2.1mm, 1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap Thermo Q Exactive HF-X Orbitrap
Ion Mode POSITIVE NEGATIVE
Units pmoles/mg tissue nanomoles/mg of tissue

Chromatography:

Chromatography ID:CH003933
Chromatography Summary:A ThermoFisher Q-Exactive HF-X mass spectrometer coupled to a Vanquish HPLC with a Waters Acquity UPLC BEH C18 (100 x 2.1 mm, 1.7 um) column was used for LC-MS/MS, with minor modifications to our previously-reported method. HPLC solvent A was 10 mM ammonium formate, 0.1% formic acid in acetonitrile:water (60:40), and solvent B was 10 mM ammonium formate, 0.1% formic acid in isopropanol:acetonitrile (90:10). A 27 min binary gradient at 0.28 mL/min was used: 0 min, 80:20 A/B; 3 min, 80:20 A/B; 5.5 min, 55:45 A/B; 8 min, 35:65 A/B; 13 min, 15:85 A/B; 14 min, 0:100 A/B; 20 min, 0:100 A/B; 20.2 min, 70:30 A/B; 27 min, 70:30 A/B.
Instrument Name:Thermo Vanquish
Column Name:Waters HPLC C18 (100 x 2.1mm, 1.7um)
Column Temperature:32
Flow Gradient:0 min, 80:20 A/B; 3 min, 80:20 A/B; 5.5 min, 55:45 A/B; 8 min, 35:65 A/B; 13 min, 15:85 A/B; 14 min, 0:100 A/B; 20 min, 0:100 A/B; 20.2 min, 70:30 A/B; 27 min, 70:30 A/B
Flow Rate:0.28 mL/min
Solvent A:60% acetonitrile/40% water; 10 mM ammonium formate; 0.1% formic acid
Solvent B:90% isopropanol/10% acetonitrile; 10 mM ammonium formate; 0.1% formic acid
Chromatography Type:Reversed phase
  logo