Summary of Study ST000443
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000330. The data can be accessed directly via it's Project DOI: 10.21228/M8W02Q This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
Study ID | ST000443 |
Study Title | Quantitative measurements of TCA cycle metabolites in T1D poor control, good control, and controls. |
Study Type | Quantitative measurements of plasma levels of tricarboxylic acid (TCA) cycle metabolites |
Study Summary | The objective of the study was to determine whether T1D with good glycemic control have persistent abnormalities of metabolites and pathways that exist in T1D with poor glycemic control. |
Institute | Mayo Clinic |
Department | Endocrinology |
Laboratory | Mayo Clinic Metabolomics Resource Core |
Last Name | Nair |
First Name | Sreekumaran |
Address | 200 First Street SW, Rochester, MN 55905 |
Nair.K@mayo.edu | |
Phone | 507-285-2415 |
Submit Date | 2016-07-20 |
Raw Data File Type(s) | d |
Analysis Type Detail | GC-MS |
Release Date | 2016-09-23 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000694 |
---|---|
Analysis type | MS |
Chromatography type | GC |
Chromatography system | Agilent 6890 |
Column | None |
MS Type | EI |
MS instrument type | Single quadrupole |
MS instrument name | Agilent 5973 |
Ion Mode | POSITIVE |
Units | micromolar |
Chromatography:
Chromatography ID: | CH000502 |
Chromatography Summary: | The derivatized extracts were analyzed with an Agilent 6890 gas chromatograph coupled with an Agilent 5973 mass selective detector. The 1-μL aliquots of the extracts were injected into a DB5-MS capillary column (30 m × 250 μm i.d., 0.25-μm film thickness; J&W Scientific, Folson, CA) using PTV injection (Gerstel CIS4 injector) in the splitless mode. The temperature of the PTV was 70 °C during injection, and 0.6 min after injection, the temperature was raised to 300 °C at a rate of 2 °C/s and held at 300 °C for 20 min. The initial GC oven temperature was 70 °C, 5 min after injection the GC oven temperature was increased with 5 °C/min to 320 °C and held for 5 min at 320 °C. Helium was used as a carrier gas and pressure programmed such that the helium flow was kept constant at a flow rate of 1.7 mL/min. Detection was achieved using MS detection in electron impact mode and full scan monitoring mode (m/z 15−800). The temperature of the ion source was set at 250 °C and that of the quadrupole at 200 °C. |
Instrument Name: | Agilent 6890 |
Column Name: | None |
Chromatography Type: | GC |