Summary of Study ST002221
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001418. The data can be accessed directly via it's Project DOI: 10.21228/M8F409 This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002221 |
Study Title | Glutaminolysis contribution to the carbon backbone of aspartate through ATP Citrate Lyase (ACLY) in ccRCC |
Study Summary | The objective of this experiment is to test the contribution of the carbons from glutamine to generation of aspartate via ATP citrate lyase (ACLY) in human epithelial renal cells HK2 and ccRCC cell lines 786-O and 786-M1A. To test this hypothesis, we incubated all cells with 13C5-glutamine in Plasmax media with or without a pharmacological inhibitor of ACLY. This is Part 8 of the study and the experimental number is MS58. |
Institute | CECAD Research Center |
Last Name | Yang |
First Name | Ming |
Address | Joseph-Stelzmann-Straße 26, Köln, Koeln, 50931, Germany |
ming.yang@uni-koeln.de | |
Phone | 4922147884306 |
Submit Date | 2022-07-15 |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2022-08-03 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003630 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Thermo Vanquish Horizon |
Column | SeQuant ZIC-pHILIC |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Exploris 240 |
Ion Mode | UNSPECIFIED |
Units | peak area |
Chromatography:
Chromatography ID: | CH002685 |
Chromatography Summary: | Chromatographic separation of metabolites was achieved using a Millipore Sequant ZIC-pHILIC analytical column (5 µm, 2.1 × 150 mm) equipped with a 2.1 × 20 mm guard column (both 5 mm particle size) with a binary solvent system. Solvent A was 20 mM ammonium carbonate, 0.05% ammonium hydroxide; Solvent B was acetonitrile. The column oven and autosampler tray were held at 40 °C and 4 °C, respectively. The chromatographic gradient was run at a flow rate of 0.200 mL/min as follows: 0–2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B. Samples were randomized and the injection volume was 5 µl. A pooled quality control (QC) sample was generated from an equal mixture of all individual samples and analysed interspersed at regular intervals. |
Instrument Name: | Thermo Vanquish Horizon |
Column Name: | SeQuant ZIC-pHILIC |
Column Temperature: | 40 |
Flow Gradient: | 0-2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B |
Flow Rate: | 0.200 mL/min |
Solvent A: | 100% water; 20 mM ammonium carbonate; 0.05% ammonium hydroxide |
Solvent B: | 100% acetonitrile |
Chromatography Type: | HILIC |