Summary of Study ST002709
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001679. The data can be accessed directly via it's Project DOI: 10.21228/M8P71W This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002709 |
Study Title | FH variant pathogenicity promotes purine salvage pathway dependence in kidney cancer |
Study Summary | The tricarboxylic citric acid cycle enzyme fumarate hydratase (FH) is a tumor suppressor. When lost in cells, its substrate fumarate accumulates to mM levels and drives oncogenic signaling and transformation. Germline alterations lead to an autosomal dominant condition known as hereditary leiomyomatosis and renal cell cancer (HLRCC) where patients are predisposed to various benign tumors and an aggressive form of kidney cancer. FH alterations of unclear significance are frequently observed with germline testing; thus, there is an unmet need to classify FH variants by their cancer-associated risk, allowing for screening, early diagnosis and treatment. Here we quantify catalytic efficiency of 74 FH variants of uncertain significance. Over half were enzymatically inactive which is strong evidence of pathogenicity. We generated a panel of HLRCC cell lines expressing FH variants with a range of catalytic activities, then correlated fumarate levels with metabolic features. We found that fumarate accumulation blocks purine biosynthesis, rendering FH-deficient cells more sensitive to the purine salvage pathway inhibitor 6-mercaptopurine. Together, these findings suggest pathogenicity of many patientassociated FH variants and reveal nucleotide salvage as a targetable vulnerability in FHdeficient cancer cells. |
Institute | University of California, Los Angeles |
Department | Biological Chemistry |
Laboratory | Heather Christofk |
Last Name | Matulionis |
First Name | Nedas |
Address | 615 Charles E Young Dr S, BSRB 354-05 |
nmatulionis@mednet.ucla.edu | |
Phone | (310) 206-0163 |
Submit Date | 2023-05-15 |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2023-06-12 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004391 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Thermo Vanquish |
Column | Merck SeQuant ZIC-pHILIC (150 x 2.1mm,5um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap |
Ion Mode | UNSPECIFIED |
Units | Peak Area |
Chromatography:
Chromatography ID: | CH003294 |
Chromatography Summary: | Dried metabolites were reconstituted in 100 µL of a 50% acetonitrile (ACN) 50% dH20 solution. Samples were vortexed and spun down for 10 min at 17,000g. 70 µL of the supernatant was then transferred to HPLC glass vials. 10 µL of these metabolite solutions were injected per analysis. Samples were run on a Vanquish (Thermo Scientific) UHPLC system with mobile phase A (20mM ammonium carbonate, pH 9.7) and mobile phase B (100% ACN) at a flow rate of 150 µL/min on a SeQuant ZIC-pHILIC Polymeric column (2.1 × 150 mm 5 μm, EMD Millipore) at 35°C. Separation was achieved with a linear gradient from 20% A to 80% A in 20 min followed by a linear gradient from 80% A to 20% A from 20 min to 20.5 min. 20% A was then held from 20.5 min to 28 min. |
Instrument Name: | Thermo Vanquish |
Column Name: | Merck SeQuant ZIC-pHILIC (150 x 2.1mm,5um) |
Column Temperature: | 35°C |
Flow Gradient: | Linear gradient from 20% A to 80% A in 20 min followed by a linear gradient from 80% A to 20% A from 20 min to 20.5 min. 20% A was then held from 20.5 min to 28 min. |
Flow Rate: | 150 µL/min |
Solvent A: | 20 mM Ammonium carbonate, pH 9.7 |
Solvent B: | 100% Acetonitrile |
Chromatography Type: | HILIC |