Summary of Study ST003021
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001773. The data can be accessed directly via it's Project DOI: 10.21228/M8J420 This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST003021 |
Study Title | Porphyrin metabolite levels in K562 cells following PKC inhibition in 2,000 nM or 100 nM FA conditions |
Study Summary | Culture of K562 cells in RPMI media containing 2000 nM or 100 nM folic acid for 6 days in the presence or absence of the PKC inhibitor, GF109203X. These samples were followed up by metabolomics targeting porphyrin metabolites. |
Institute | Boston Children's Hospital, Harvard Medical School |
Department | pathology |
Laboratory | Kanarek Lab |
Last Name | Kanarek |
First Name | Naama |
Address | Enders 1116.2, 300 Longwood Ave, Boston, MA 02115 |
naama.kanarek@childrens.harvard.edu | |
Phone | 6173557433 |
Submit Date | 2023-12-19 |
Num Groups | 5 |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2024-02-13 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004955 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Vanquish |
Column | Phenomenex Kinetex C18 (150 x 3mm,2.6um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE |
Units | Arbitrary Units |
Chromatography:
Chromatography ID: | CH003739 |
Chromatography Summary: | Sample (5 uL) was injected onto a 2.6 μm, 150 x 3 mm C18 column (Phenomenex, 00F-4462-Y0) equipped with a 3.0 mm safe-guard column (Phenomenex, AJ0-8775). Column compartment was heated to 45 ºC. Porphyrins were separated with a chromatographic gradient at a flow rate of 0.800 ml min−1 as follows: 0–2 min: 5% B; 2-19min: linear gradient from 5% to 95% B; 19-21min: 95% B; 21.1-23min: return to 5% B. |
Instrument Name: | Thermo Vanquish |
Column Name: | Phenomenex Kinetex C18 (150 x 3mm,2.6um) |
Column Temperature: | 45 |
Flow Gradient: | 0–2 min: 5% B; 2-19min: linear gradient from 5% to 95% B; 19-21min: 95% B; 21.1-23min: return to 5% B. |
Flow Rate: | 0.8 mL/min |
Solvent A: | 95% water/5% acetonitrile; 0.1% formic acid |
Solvent B: | 5% water/95% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |