MB Sample ID: SA005530
Local Sample ID: | 9PHS-5 |
Subject ID: | SU000119 |
Subject Type: | Invertebrate |
Subject Species: | Caenorhabditis elegans |
Taxonomy ID: | 6239 |
Genotype Strain: | N2 |
Age Or Age Range: | Young adult |
Gender: | Hermaphrodite |
Species Group: | Invertebrate |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000165 | AN000166 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Thermo Dionex Ultimate 3000 RS | Thermo Dionex Ultimate 3000 RS |
Column | Waters Acquity HSS T3 (100 x 2.1mm,1.8um) | Waters Acquity HSS T3 (100 x 2.1mm,1.8um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units |
MS:
MS ID: | MS000134 |
Analysis ID: | AN000165 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The Q-Exactive was equipped with a Heated Electrospray Ionization (HESI) source which operated at a spray temperature of 350 ?C, a spray voltage of 3.5 kV, and sheath and auxiliary gas flow rates of 50 and 10 arbitrary units, respectively. The capillary temperature was held at 325 ?C, and the S-lens RF Level was set to 40%. The FR resolution was set to 70,000 at m/z 200. The accuracy achieved was routinely less than 1.5 ppm. |
Ion Mode: | POSITIVE |
Capillary Temperature: | 325 ?C |
Ion Source Temperature: | 350 ?C |
Ion Spray Voltage: | 3.5 kV |
Resolution Setting: | 70000 |
MS ID: | MS000135 |
Analysis ID: | AN000166 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The Q-Exactive was equipped with a Heated Electrospray Ionization (HESI) source which operated at a spray temperature of 350 ?C, a spray voltage of 3.5 kV, and sheath and auxiliary gas flow rates of 50 and 10 arbitrary units, respectively. The capillary temperature was held at 325 ?C, and the S-lens RF Level was set to 40%. The FR resolution was set to 70,000 at m/z 200. The accuracy achieved was routinely less than 1.5 ppm. |
Ion Mode: | NEGATIVE |
Capillary Temperature: | 325 ?C |
Ion Source Temperature: | 350 ?C |
Ion Spray Voltage: | 3.5 kV |
Resolution Setting: | 70000 |