Return to study ST001185 main page
MB Sample ID: SA082244
Local Sample ID: | R6 |
Subject ID: | SU001251 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN001967 | AN001968 | AN001969 |
---|---|---|---|
Analysis type | MS | MS | MS |
Chromatography type | Reversed phase | Reversed phase | Reversed phase |
Chromatography system | Thermo Vanquish | Thermo Vanquish | Thermo Vanquish |
Column | Thermo Fisher Accucore C18 (100 x 2.1mm,5um) | Thermo Fisher Accucore C18 (100 x 2.1mm,5um) | Phenomenex Kinetex C18+ (100 x 2.1mm,1.7um) |
MS Type | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE | NEGATIVE | UNSPECIFIED |
Units | Peak area normalized to internal standard | Peak area normalized to internal standard | Peak Area |
MS:
MS ID: | MS001821 |
Analysis ID: | AN001967 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | LIPIDOMICS: Samples were analyzed using full scan accurate mass at a resolution of 70K in positive and negative mode and 17.5K for ddMS2. Thermo Fisher LipidSearch 4.2.2 software was used for peak quantification, alignment, and MS2 identification. Peak areas were normalized to internal standard. |
Ion Mode: | POSITIVE |
MS ID: | MS001822 |
Analysis ID: | AN001968 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | LIPIDOMICS: Samples were analyzed using full scan accurate mass at a resolution of 70K in positive and negative mode and 17.5K for ddMS2. Thermo Fisher LipidSearch 4.2.2 software was used for peak quantification, alignment, and MS2 identification. Peak areas were normalized to internal standard. |
Ion Mode: | NEGATIVE |
MS ID: | MS001823 |
Analysis ID: | AN001969 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | METABOLOMICS: Metabolites were identified in both positive and negative mode by accurate mass (≤ 10 ppm) and retention time. Integrated peak areas were then extracted manually using Quan Browser (Thermo Fisher Xcalibur ver 2.7) and normalized to internal standard. |
Ion Mode: | UNSPECIFIED |