MB Sample ID: SA086141

Local Sample ID:Plasma_59
Subject ID:SU001284
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002029 AN002030
Analysis type MS MS
Chromatography type Reversed phase HILIC
Chromatography system Agilent 6545 Agilent 6520
Column Agilent Zorbax RRHD SB-C18 (100 x 2.1mm,1.8um) Phenomenex Kinetex HILIC 100A (50 x 2.1 mm,2.6um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6545 QTOF Agilent 6520 QTOF
Ion Mode POSITIVE POSITIVE
Units Abundance (Log2) Abundance (Log2)

MS:

MS ID:MS001882
Analysis ID:AN002029
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:Agilent 6545 Quadrupole Time-of-Flight mass spectrometer (QTOF-MS) in positive ionization mode with dual AJS ESI source, mass range 50–1700 m/z, scan rate 2.00, gas temperature 300 °C, gas flow 12.0 L/min, nebulizer 35 psi, sheath gas temperature 275°C, skimmer 65 V, capillary voltage 3500 V, fragmentor 120 V, reference masses 121.050873 and 922.009798 (Agilent reference mix). The analysis was replicated for tandem MS of selected compounds using a scan range 50–1700m/z, and 10, 20, and 40 eV collision energies with a 500 ms/spectra acquisition time, 1.3 m/z (narrow) isolation width, and 0.25 minute delta retention time.
Ion Mode:POSITIVE
  
MS ID:MS001883
Analysis ID:AN002030
Instrument Name:Agilent 6520 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:Agilent 6520 QTOF-MS in positive ionization mode with dual ESI source, mass range 50–1700 m/z, scan rate 2.00, gas temperature 325 °C, gas flow 12.0 L/min, nebulizer 30 psi, skimmer 60 V, capillary voltage 4000 V, fragmentor 120 V, reference masses 121.050873 and 922.009798 (Agilent reference mix). The analysis was replicated for tandem MS of selected compounds using a scan range 50–1700m/z, and 10, 20, and 40 eV collision energies with a 500 ms/spectra acquisition time, 1.3 m/z (narrow) isolation width, and 0.25 minute delta retention time.
Ion Mode:POSITIVE