Return to study ST001365 main page

MB Sample ID: SA099252

Local Sample ID:Sample_15137
Subject ID:SU001439
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:21-56
Gender:Male and female
Human Inclusion Criteria:Cystic Fibrosis, adults

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002272
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity I-Class
Column Waters Acquity BEH C18 (50 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name Waters Quattro Premier XE
Ion Mode POSITIVE
Units micromolar

MS:

MS ID:MS002116
Analysis ID:AN002272
Instrument Name:Waters Quattro Premier XE
Instrument Type:Triple quadrupole
MS Type:ESI
MS Comments:The retention times (RT) and MS/MS parameters for reach antibiotic were determined using the QuanOpt function in the Waters Masslynx software. Peaks were automatically picked and filtered in QuanLynx. Downstream filtering was conducted in R. Data were imported into R and the following criteria for filtering were applied: minimum peak area under the curve of 20 and signal to noise ratio of 10. Because several of the antibiotics had high carry-over rates (including ciprofloxacin, levofloxacin, and trimethoprim), sample peaks were also filtered out when the AUC was lower than the wash ran directly before the sample. Scripts can be found at the following github link: https://github.com/tgallagh/LCMS_Antibiotics.
Ion Mode:POSITIVE
Capillary Voltage:3.3 kV
Collision Gas:Nitrogen
Source Temperature:125
Desolvation Gas Flow:800 L/h
Desolvation Temperature:400
Acquisition Parameters File:CF_antibiotics.exp
Analysis Protocol File:CF_antibiotics_quant.mdb